Performance of thin Vivacit-E hip liners: no relevant wear during ex vivo testing at high acetabular inclination angle.

Saikko Vesa, Neuvonen Perttu, Reito Aleksi

СтатьяПеревод в очереди

Аннотация

BACKGROUND AND PURPOSE: There is concern among clinicians regarding the performance of thin, highly cross-linked polyethylene acetabular liners at high inclination angles that cause edge contact and high contact stresses. We studied ex vivo wear performance of thin, vitamin-E grafted, highly cross-linked polyethylene (Vivacit-E) liners in relation to high acetabular inclination angle.

MATERIALS AND

METHODS

Wear of Vivacit-E acetabular liners (thickness 4.0-5.7 mm) was studied with a validated hip joint simulator at 2 different acetabular inclination angles, 40° (optimal) and 65° (high). The test simulated walking. Wear was evaluated gravimetrically and dimensionally.

RESULTS

At the optimal inclination angle, slight weight gain occurred. At the high angle, the liners lost weight slightly. Due to the minimal weight loss, gravimetric wear rates were difficult to determine. Linear wear was below the detection limit of 0.01 mm.

CONCLUSION

Even with the high acetabular inclination angle, the prostheses performed well in the present test conditions. The wear rates of the liners were estimated to be clearly below the osteolysis threshold of 0.05 mm/year, below which osteolysis is absent. Since the present hip joint simulator has been shown to produce clinically relevant wear, these ex vivo results are likely to reflect the in vivo behavior of the design.

Для цитирования: Saikko Vesa, Neuvonen Perttu, Reito Aleksi. Performance of thin Vivacit-E hip liners: no relevant wear during ex vivo testing at high acetabular inclination angle.. Acta orthopaedica. 2022. DOI: 10.2340/17453674.2022.5359.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.