Prospective Trial of Sacroiliac Joint Fusion Using 3D-Printed Triangular Titanium Implants: 24-Month Follow-Up.

Patel Vikas, Kovalsky Don, Meyer S Craig, Chowdhary Abhineet, LaCombe Julie, Lockstadt Harry, Techy Fernando, Langel Casey, Limoni Robert, Yuan Philip, Kranenburg Andy, Cher Daniel, Tender Gabriel

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Strong evidence supports minimally invasive sacroiliac joint (SIJ) fusion using triangular titanium implants (TTI) for chronic SIJ dysfunction.

OBJECTIVE

To report safety and effectiveness of SIJF using a 3D-printed TTI at 24 months.

METHODS

SIJF with TTI was performed in 51 subjects. Structured follow-up occurred at 3, 6, 12 and 24 months. Both quality of life questionnaires and functional tests were performed at all study visits.

RESULTS

84% of subjects were available for 24-month follow-up. Observed were rapid and persistent improvements in dysfunction due to pain (Oswestry Disability Index [ODI], mean 52.8 at baseline and 28.3 at 24 months, p<0.0001) and SIJ pain ratings (mean 78.5 at baseline [0-100 scale] to 21.5 at 24 months). Opioid use for SIJ pain decreased markedly from baseline. Physical function tests impaired by SIJ pain showed persistent improvements compared to baseline. There was no evidence of device breakage, migration or subsidence and few late adverse events occurred attributable to the device.

CONCLUSION

In this prospective study, SIJF using 3D-printed TTI resulted in immediate, marked and persistent improvements in pain and quality of life, with improved physical function, reduced opioid use and a low rate of late device-related adverse events.

LEVEL OF EVIDENCE: Level II.

Для цитирования: Patel Vikas, Kovalsky Don, Meyer S Craig, Chowdhary Abhineet, LaCombe Julie, Lockstadt Harry, Techy Fernando, Langel Casey, Limoni Robert, Yuan Philip, Kranenburg Andy, Cher Daniel, Tender Gabriel. Prospective Trial of Sacroiliac Joint Fusion Using 3D-Printed Triangular Titanium Implants: 24-Month Follow-Up.. Medical devices (Auckland, N.Z.). 2021. DOI: 10.2147/mder.s314828.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.