Archives of orthopaedic and trauma surgery · 2021
Handheld computer-navigated constrained total knee arthroplasty for complex extra-articular deformities.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Pietsch M, Hochegger M, Djahani O, Mlaker G, Eder-Halbedl M, Hofstädter Th
Аннотация
INTRODUCTION: The present study aimed to assess the postoperative alignment and clinical outcomes of patients with complex extra-articular deformities (EADs) undergoing computer-assisted surgery (CAS) for constrained total knee arthroplasty (TKA) with modular stem extensions.
MATERIALS AND METHODS: From May 2015 to July 2018, ten patients with EADs scheduled for constrained TKA were enrolled retrospectively. The preoperative average deviation from neutral (= 180°) mechanical axis was 15.3° (range of coronal alignment: 150.9° varus-202.9° valgus). Alignment was assessed using an accelerometer-based handheld CAS system. On long-leg films, the positions of the components and possible stems were analysed and templated preoperatively. The average follow-up was 3.3 years (range: 2.0-4.6 years).
RESULTS: The postoperative mechanical axis was within ± 3.0° from neutral in nine patients. In all patients, the Knee Society score (KSS) and range of motion improved significantly. A constrained condylar and a rotating hinge prosthesis were used in five patients each. In eight patients, the 100-mm cementless stem that was preferred by the authors was found to be unusable for the femur or the tibia in the planning stage. For the femur, a cementless 100-mm stem was used in three, and a cemented 30-mm stem in five patients; a femoral stem was not usable in two patients. For the tibia, a cementless 100-mm stem was used in six, and a cemented 30-mm stem in two patients; a monoblock rotating hinge tibia was used in two patients.
CONCLUSIONS: Complex EADs were excellently managed during constrained TKA implantation using the handheld CAS system. Templating allowed the possible stem lengths to be identified and prevented anatomical conflict with the CAS-configured mechanical alignment. Limb alignment and function improved significantly after surgery. No intra- or postoperative complications occurred.
LEVEL OF EVIDENCE: Level IV, retrospective study.
MATERIALS AND METHODS: From May 2015 to July 2018, ten patients with EADs scheduled for constrained TKA were enrolled retrospectively. The preoperative average deviation from neutral (= 180°) mechanical axis was 15.3° (range of coronal alignment: 150.9° varus-202.9° valgus). Alignment was assessed using an accelerometer-based handheld CAS system. On long-leg films, the positions of the components and possible stems were analysed and templated preoperatively. The average follow-up was 3.3 years (range: 2.0-4.6 years).
RESULTS: The postoperative mechanical axis was within ± 3.0° from neutral in nine patients. In all patients, the Knee Society score (KSS) and range of motion improved significantly. A constrained condylar and a rotating hinge prosthesis were used in five patients each. In eight patients, the 100-mm cementless stem that was preferred by the authors was found to be unusable for the femur or the tibia in the planning stage. For the femur, a cementless 100-mm stem was used in three, and a cemented 30-mm stem in five patients; a femoral stem was not usable in two patients. For the tibia, a cementless 100-mm stem was used in six, and a cemented 30-mm stem in two patients; a monoblock rotating hinge tibia was used in two patients.
CONCLUSIONS: Complex EADs were excellently managed during constrained TKA implantation using the handheld CAS system. Templating allowed the possible stem lengths to be identified and prevented anatomical conflict with the CAS-configured mechanical alignment. Limb alignment and function improved significantly after surgery. No intra- or postoperative complications occurred.
LEVEL OF EVIDENCE: Level IV, retrospective study.