Clinical results at 10 years of minimum follow-up with the ABG 2 hip arthroplasty, matched with ceramic-on-ceramic bearings.

Coulomb Remy, Mansour Jad, Essig Jérome, Asencio Gérard, Kouyoumdjian Pascal

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

The current study aimed as a primary goal is to assess the results of a ceramic-on-ceramic (CoC) bearing hip system matched with ABG (Anatomic Benoist Girard) 2 components in terms of survivorship. Secondary objectives addressed specifically ceramic-related complications as well as specific patterns at the bone-implant interface.

MATERIAL AND

METHODS

This is a retrospective bicentric continuous series involving 147 patients (95 males vs. 52 females) who underwent ABG 2 arthroplasties with CoC bearings. One hundred and twenty-five hips were closely followed-up at a mean period of 11.3 years.

RESULTS AND DISCUSSION: With a mean follow-up of 11.3 years, nine cases (5.7%) underwent revision surgery, four caused by acetabular aseptic loosening, three by deep infections, one ceramic head fracture, and one femoro-acetabular impingement. The global survivorship was 92.2% at 12.7 years. The Harris Hip Score (HHS) mean scores increased post-operatively from 50.1 up to 96.1 points (p < 0.001). All stems featured patterns of radiological osseous integration onto the hydroxyapatite (HA)-coated zones. No radiological wear or osteolysis of ceramic bearings was demonstrated however, five patients reported hip squeaking using this bearing. This study demonstrated excellent results at mid-term follow-up in patients younger than 70 years of age using cementless ABG 2 components coupled with CoC bearings with no increase in complication rate.

Для цитирования: Coulomb Remy, Mansour Jad, Essig Jérome, Asencio Gérard, Kouyoumdjian Pascal. Clinical results at 10 years of minimum follow-up with the ABG 2 hip arthroplasty, matched with ceramic-on-ceramic bearings.. SICOT-J. 2022. DOI: 10.1051/sicotj/2022032.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.