DOI 10.1080/10255842.2025.2530638

Медицинская библиотека · Computer methods in biomechanics and biomedical engineering · 2025

Finite element analysis of biomechanical effects of continuous versus interval pedicle screw configurations in scoliosis correction and optimization of dual-geometry screw design.

He Chunshan, Dou Shixin, Ma Xiaoying, Hou Zhenhua

СтатьяПеревод в очереди

Аннотация

PURPOSE: To optimize scoliosis correction strategies by comparing continuous and interval pedicle screw configurations and proposing a dual-geometry screw design.

METHODS

A patient-specific T11-L5 scoliotic spine model was reconstructed via finite element analysis (FEA). Continuous and interval screw placements were evaluated for biomechanical performance. A novel dual-geometry screw (tapered-cylindrical transition) was developed.

RESULTS

Continuous configurations achieved a 43.5% reduction in displacement (1.33 mm vs. 2.36 mm) and a 29.7% decrease in screw stress (444.08 MPa vs. 631.35 MPa). The dual-geometry screw lowered drilling stress (16.5%, p < 0.05) and displacement heterogeneity (22.4%).

CONCLUSION

Continuous screws enhance stability through synergistic load transfer, while dual-geometry screws mitigate interfacial damage. This provides biomechanical criteria for clinical scoliosis correction.

Для цитирования: He Chunshan, Dou Shixin, Ma Xiaoying, Hou Zhenhua. Finite element analysis of biomechanical effects of continuous versus interval pedicle screw configurations in scoliosis correction and optimization of dual-geometry screw design.. Computer methods in biomechanics and biomedical engineering. 2025. DOI: 10.1080/10255842.2025.2530638.