A multicenter study of shock pathways for subcutaneous implantable defibrillators.

Kuschyk Jürgen, Milasinovic Goran, Kühlkamp Volker, Roberts Paul R, Zabel Markus, Molin Franck, Shorofsky Stephen, Stromberg Kurt D, DeGroot Paul J, Murgatroyd Francis D, Solo Study Investigators

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

INTRODUCTION

A purely subcutaneous implantable cardioverter defibrillator (ICD) requires higher energy but may be an effective alternative to transvenous ICDs to deliver lifesaving therapies.

OBJECTIVE

To identify combinations of anteroposterior subcutaneous shock pathways and waveforms with defibrillation efficacy comparable to transvenous ICDs.

METHODS

Defibrillation testing was performed in 141 patients temporarily implanted with an active can emulator and subcutaneous coil electrodes. The patients were subdivided into 5 groups within 2 study phases. In all groups, a posterior electrode was positioned with its tip close to the spine. In the first study phase, 2 different can locations were evaluated: (1) an inframammary pocket (IM-1-750), or (2) a conventional infraclavicular pocket (IC-1-750). In both cases, a 70 J biphasic shock was used (peak voltage 750 V; 270 μF capacitance). In the second phase, configuration IC-1-750 was enhanced by the addition of a second (parasternal) subcutaneous electrode (IC-2-750). Furthermore, the effects of a different 70 J shock waveform (1,000 V, 160 μF) were evaluated for configurations IM-1-750 and IC-2-750 (becoming IM-1-1000 and IC-2-1000).

RESULTS

The proportion of patients satisfying a defibrillation safety margin test of 2 consecutive successes at ≤50 J was 74%, 11%, and 44%, respectively, for the IM-1-750, IC-1-750, and IC-2-750 configurations, and 93% and 86% for the IM-1-1000 and IC-2-1000 configurations.

CONCLUSIONS

Defibrillation efficacy comparable to that of a transvenous system was achieved with an anteroposterior subcutaneous ICD configuration, with 160 μF capacitance, 1,000 V, and 70 J output. An infraclavicular pocket location becomes feasible if a parasternal subcutaneous coil is added.

Для цитирования: Kuschyk Jürgen, Milasinovic Goran, Kühlkamp Volker, Roberts Paul R, Zabel Markus, Molin Franck, Shorofsky Stephen, Stromberg Kurt D, DeGroot Paul J, Murgatroyd Francis D, Solo Study Investigators. A multicenter study of shock pathways for subcutaneous implantable defibrillators.. Journal of cardiovascular electrophysiology. 2013. DOI: 10.1111/jce.12281.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.