MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/10225536221109960

Медицинская библиотека · Journal of orthopaedic surgery (Hong Kong) · 2022

A multi-center retrospective comparative study of third generation Ceramic-on- Ceramic total hip arthroplasty in patients younger than 45 years with or without the sandwich liner: A ten-year minimum.

He Bangjian, Li Xigong, Dong Rui, Tong Peijian, Sun Junying

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Ceramic-on-ceramic couplings are attractive alternative bearing surfaces that have been reported to eliminate or reduce problems related to polyethylene wear debris. However, the material in THA still remains one of the major concerns about the risk of fracture, due to its brittleness.

OBJECTIVE

The present study aims at reporting the fracture rate of a series of ceramic-on-ceramic THAs with use of the sandwich liner combined with a ceramic femoral head, and attempt to detect the relative risk factors, possible cause and assesse the medium-term clinical results.

METHODS

We retrospectively evaluated 282 patients (300 hips) with use of the sandwich liner ceramic-on-ceramic THA between 2001 and 2009 at three-centers. Patient assessment was based on demographic factors, including age, weight, gender and body-mass index. All patients were evaluated clinically and radio-graphically or computed tomography in consideration of dislocation, osteolysis, periprosthetic fracture, infection, loosening and implant fracture.

RESULTS

five ceramic sandwich liners fracture (1.7%) were observed at an average of 7.3 years follow-up. These factors were irrelevant to the ceramic liner fracture, including age (p = 0.205), weight (p = 0.241), gender (p = 0.553), body-mass index (p = 0.736), inclination (p = 0.727), and anteversion (p = 0.606). The overall survival was 91.4% at 12 years with revision as the endpoint. Other complications included dislocation in two, perprosthetic fracture in two and osteolysis in eight hips. No hip had aseptic loosening of the implants was seen.

CONCLUSIONS

We found that the sandwich liner may be lead to a high rate of alumina fracture and osteolysis. We have discontinued the use of sandwich liner with THA since 2009.

Для цитирования: He Bangjian, Li Xigong, Dong Rui, Tong Peijian, Sun Junying. A multi-center retrospective comparative study of third generation Ceramic-on- Ceramic total hip arthroplasty in patients younger than 45 years with or without the sandwich liner: A ten-year minimum.. Journal of orthopaedic surgery (Hong Kong). 2022. DOI: 10.1177/10225536221109960.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.