MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.arth.2016.05.065

Медицинская библиотека · The Journal of arthroplasty · 2016

Determining False Positive Rates of Leukocyte Esterase Reagent Strip When Used as a Detection Tool for Joint Infection.

McNabb David Clinton, Dennis Douglas A, Kim Raymond H, Miner Todd M, Yang Charlie C, Jennings Jason M

Клиническое исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Use of leukocyte esterase (LE) testing of synovial fluid as an adjunct to the infection workup in total joint arthroplasty patients has been advocated. The purpose of this study was to determine the false positive rate of this test.

METHODS

Two hundred patients with the diagnosis of degenerative osteoarthritis undergoing a total knee arthroplasty were identified for this study. The knee joint was aspirated under sterile conditions before performing the total knee arthroplasty. The fluid was analyzed with an LE reagent strip.

RESULTS

There were 27 bloody and 17 dry aspirations. One hundred forty-nine patients produced an aspiration that allowed for LE testing. There was 1 positive LE result. The specificity of the LE test was found to be 99.3%.

CONCLUSION

These data suggest the LE strip as a part of the workup for infection in a native knee should yield few false positive results.

Для цитирования: McNabb David Clinton, Dennis Douglas A, Kim Raymond H, Miner Todd M, Yang Charlie C, Jennings Jason M. Determining False Positive Rates of Leukocyte Esterase Reagent Strip When Used as a Detection Tool for Joint Infection.. The Journal of arthroplasty. 2016. DOI: 10.1016/j.arth.2016.05.065.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.