Comparative FE biomechanical and microbial adhesion analyses on an implanted humerus.
Аннотация
OBJECTIVE
In this study a multi fragment humeral fracture, treated with locking plate system implant, was investigate and compared with a healthy humerus by the mining of a Finite Element (FE) analysis. Locking plate implant, in this case AxSOS 3® Titanium produced by Stryker is the preferred solution in presence of multiple fracture or osteoporosis.
METHODS
Loading conditions were imposed by rotating of 52,5° respect the vertical axe, both the humeri (healthy and fractured), fixing distal end, and loading the top of bones with a vertical force of 543 N (Newton). This finite element analysis aimed to compare stability of implanted humerus, implant-bone interface, stress shielding, with those related to a healthy one. A microbial adhesion analysis was also performed on the implant's material.
RESULTS
Results obtained by FE analysis confirm a good agreement of the mechanical behavior of the models tested. The maximum values of the registered stressed, are of about 45 MPa (Mega Pascal) for the intact humerus and 113 MPa for the fractured one. Displacements, confirm higher values on the fractured humerus and no viable bacteria were found after microbial adhesion analyses.
Conclusion
comparison between healthy and fractured humerus showed an optimal stability of the implant, when contact surfaces optimization and screws insertion are correctly performed.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.