Unexpected varus deformity and concomitant metal ion release and MRI findings of modular-neck hip stems: descriptive RSA study in 75 hips with 8 years' follow-up.

Kiernan Sverrir, Kaptein Bart, Flivik Carl, Sundberg Martin, Flivik Gunnar

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

Background and purpose - Modular-neck hip stems have been identified with corrosion-related problems from the neck-stem junction. We report an ongoing varus deformity of modular-neck hip stems with simultaneous metal ion release observed during a study comparing the migration of modular vs. standard hip stems.Patients and methods - We followed 50 patients with modular and 25 with standard neck stems using radiostereometry (RSA). At 5-year follow-up, we noted a compromised integrity of the modular stem with varus deformity in the neck-stem interface. Changes in head-tip distance as well as whole-blood ion concentration and MRI findings were analyzed. The modular stems were followed further up to 8 years.Results - The head-tip distance decreased continuously by 0.15 mm per year resulting in 1.2 (95% CI 1.0-1.4) mm at 8 years for modular stems, while for the standard stems at 5 years, the decrease was 0.09 (CI 0.0-0.2) mm or 0.02 mm/year. For the modular stems, the reduction in head-tip distance correlated to the increase in whole-blood cobalt concentration at 8 years but not to the MRI grading of tissue reactions. At 5 years, cobalt levels were 4.9 µg/L for modular stems and at 8 years 4.8 µg/L, whereas for standard stems this was 1.0 µg/L. After 8 years, 9 of 72 stems had been revised for different reasons, but only 1 with obvious adverse local tissue reaction (ALTR).Interpretation - We present a surprisingly large progressive deformation at the modular neck-stem junction, but so far without a definite clinical problem. Even the femoral head seems to show slight compression onto the taper over time. A high rate of revisions for the modular type of this stem has raised general concerns, and it has been recalled from the market.

Для цитирования: Kiernan Sverrir, Kaptein Bart, Flivik Carl, Sundberg Martin, Flivik Gunnar. Unexpected varus deformity and concomitant metal ion release and MRI findings of modular-neck hip stems: descriptive RSA study in 75 hips with 8 years' follow-up.. Acta orthopaedica. 2020. DOI: 10.1080/17453674.2020.1853387.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.