MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2106/jbjs.cc.23.00043

Медицинская библиотека · JBJS case connector · 2023

Failure of a Femoral Revision Knee Component with Chemical Corrosion and Elevated Metal Ions: A Case Report and Literature Review.

Holland Christopher T, Shenoy Aarti, Lachiewicz Paul F

ОбзорКлинический случайСтатьяПеревод в очереди

Аннотация

CASE: Failure of the modular junction in revision total knee arthroplasty is a rare complication. We report a patient with late, atraumatic failure of a modern, modular revision femoral component, with preoperative elevation of serum cobalt and chromium levels. Retrieval analysis showed extensive chemical corrosion.

CONCLUSION

Failure of a modern, modular femoral component may cause metal synovitis and elevated serum metal levels. Subtle radiographic changes and preoperative serum metal levels may identify this complication.

Для цитирования: Holland Christopher T, Shenoy Aarti, Lachiewicz Paul F. Failure of a Femoral Revision Knee Component with Chemical Corrosion and Elevated Metal Ions: A Case Report and Literature Review.. JBJS case connector. 2023. DOI: 10.2106/jbjs.cc.23.00043.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.