Nielson Trent, Owens Garrett, Miller Brandon, Meneghini Ethan, Deckard Evan R, Meneghini R Michael

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Highly cross-linked polyethylene with vitamin E (VE-HXLPE) has shown superior tribological properties and has been rapidly adopted in total hip arthroplasty. However, the majority of studies compare VE-HXLPE to conventional or moderately cross-linked polyethylene using standard femoral head sizes. This study's purpose was 2-fold: (1) compare radiographic femoral head penetration (FHP) between VE-HXLPE and HXLPE and (2) evaluate FHP in large femoral heads ≥40 mm.

METHODS

One hundred forty-two consecutive primary total hip arthroplasties using ceramic femoral heads (n = 84 VE-HXLPE; n = 58 HXLPE) in a single implant system were retrospectively reviewed. FHP was measured radiographically utilizing Martell method at 4-week, 1-year, and latest radiographs. FHP, cup position, and demographic variables were compared between VE-HXLPE and HXLPE liners.

RESULTS

Median linear FHP was lower for VE-HXLPE compared to HXLPE during the initial "bedding-in" period between 4-week and 1-year (0.383 vs 0.551 mm, P = .650) and between 1-year and latest follow-up (0.131 vs 0.270 mm/y, P = .636) although without statistical significance. Acetabular cup inclination and anteversion did not influence linear or volumetric FHP (P ≥ .204). Large femoral heads (≥40 mm) were predictive of higher FHP during the early bedding-in period (P ≤ .025) but did not have an effect beyond 1 year in multivariate regression with numbers available. No radiographic osteolysis was observed in any case.

CONCLUSION

These findings support others that VE-HXLPE is the optimal polyethylene bearing surface to minimize FHP during the bedding-in period and beyond. Surprisingly, large ceramic femoral heads appear to influence FHP during the initial bedding-in period but do not increase FHP beyond 1 year. Further longer term follow-up remains warranted.

LEVEL OF EVIDENCE: III.

Для цитирования: Nielson Trent, Owens Garrett, Miller Brandon, Meneghini Ethan, Deckard Evan R, Meneghini R Michael. Large Femoral Heads in Total Hip Arthroplasty With Vitamin E Highly Cross-Linked Polyethylene: Head Penetration Rates Compared to Highly Cross-Linked Polyethylene.. The Journal of arthroplasty. 2022. DOI: 10.1016/j.arth.2022.01.075.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.