Unexpected Wear of a Uniquely Designed Moderately Cross-Linked Polyethylene in Total Hip Arthroplasty.

Yakkanti Ramakanth R, Ocksrider Justin L, Patel Anand A, Kolevar Matthew P, Moore Rebecca D, Rimnac Clare M, Kraay Matthew J, Wright Timothy M, Baral Elexis C, Robinson Raymond P

СтатьяПеревод в очереди

Аннотация

BACKGROUND

A uniquely designed, non-heat-treated moderately cross-linked acetabular polyethylene liner used in total hip arthroplasty (THA) demonstrated excessive wear during routine follow-up, prompting an evaluation of the linear wear rate.

METHODS

All THAs were performed by the senior author. The study group included 38 THAs using the uniquely designed polyethylene in question, compared to a control group of 21 THAs using another moderately cross-linked polyethylene with good 10-year outcomes. Two-dimensional linear head penetration wear measurements were obtained using the Martell Hip Analysis Suite, and retrieval analysis was performed on two liners.

RESULTS

The study group had a significantly higher average penetration rate of 0.089 mm/y than the control group average rate of 0.047 mm/y (P = .04). Forty-five percent of the study group had a wear rate above the osteolysis threshold (0.1 mm/y), compared to 24% in the control group. Macroscopic analysis of two retrieved liners validated the radiographic findings.

CONCLUSION

The data suggest unexpectedly higher wear rates for a moderately cross-linked polyethylene design, with nearly half of the study group at risk for osteolysis. Further registry or database analyses of this particular moderately cross-linked polyethylene are warranted.

Для цитирования: Yakkanti Ramakanth R, Ocksrider Justin L, Patel Anand A, Kolevar Matthew P, Moore Rebecca D, Rimnac Clare M, Kraay Matthew J, Wright Timothy M, Baral Elexis C, Robinson Raymond P. Unexpected Wear of a Uniquely Designed Moderately Cross-Linked Polyethylene in Total Hip Arthroplasty.. The Journal of arthroplasty. 2022. DOI: 10.1016/j.arth.2022.01.093.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.