DOI 10.1016/j.knee.2021.10.005
Highly cross-linked polyethylene (HXLPE) is equivalent to conventional polyethylene (CPE) in total knee arthroplasty: A systematic review and meta-analysis.
Аннотация
BACKGROUND
The benefits of HXLPE in total knee arthroplasty (TKA) have not been as evident as total hip arthroplasty (THA). A systematic review and meta-analysis to assess the impact of highly-crosslinked polyethylene (HXLPE) on TKA outcomes compared to conventional polyethylene (CPE) is described.
METHODS
All studies comparing HXLPE with CPE for primary TKA were included for analysis. The minimum dataset included revision rates, indication for revision, aseptic component loosening and follow-up time. The primary outcome variables were all-cause revision, aseptic revision, revision for loosening, radiographic component loosening, osteolysis and incidence of radiolucent lines. Secondary outcome measures included postoperative functional knee scores. A random-effects meta-analysis allowing for all missing data was performed for all primary outcome variables.
RESULTS
Six studies met the inclusion criteria. In total, there were 2,234 knees (1,105 HXLPE and 1,129 CPE). The combined mean follow-up for all studies was 6 years. The aseptic revision rate in the HXLPE group was 1.02% compared to 1.97% in the CPE group. There was no difference in the rate of all-cause revision (p = 0.131), aseptic revision (p = 0.298) or revision for component loosening (p = 0.206) between the two groups. Radiographic loosening (p = 0.200), radiolucent lines (p = 0.123) and osteolysis (p = 0.604) was similar between both groups. Functional outcomes were similar between groups.
CONCLUSION
The use of HXLPE in TKA yields similar results for clinical and radiographic outcomes when compared to CPE at midterm follow-up. HXLPE does not confer the same advantages to TKA as seen in THA.