MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/mus.27127

Медицинская библиотека · Muscle & nerve · 2020

Proximal fibular nerve conduction studies to tibialis anterior: Optimal E2 (reference electrode) placement.

Day Timothy J

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Several E2 (reference electrode) positions are described for fibular (peroneal) nerve conduction studies to tibialis anterior (TA).

METHODS

This study compared the contribution of different E2 sites to the TA motor response, using remote referential recordings and different bipolar montages.

RESULTS

The medial knee contributes minimal electrical activity to the bipolar TA recordings, whereas tibial, ankle, and toe references resulted in very similar, moderate amplitude contributions consistent with far field potentials. These observations were very similar in controls and in patients with lower leg symptoms and signs.

CONCLUSIONS

Standard montages using distal leg or foot E2 sites result in lower amplitudes with distortion arising from the E2 electrode, compared with the TA-Knee montage. Optimal measurement of the TA motor response is achieved using a medial knee reference, without compromising measures of fibular nerve conduction across the knee.

Для цитирования: Day Timothy J. Proximal fibular nerve conduction studies to tibialis anterior: Optimal E2 (reference electrode) placement.. Muscle & nerve. 2020. DOI: 10.1002/mus.27127.