DOI 10.1016/j.jhsa.2025.03.009

Медицинская библиотека · The Journal of hand surgery · 2025

Unicortical Versus Bicortical Screws for Dorsal Plate Fixation of Phalangeal Fractures.

Sivakumar Brahman Shankar, Bissell Lianne, Hile Mark, Clarke Elizabeth

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

PURPOSE: To compare unicortical and bicortical screw fixation for dorsal plating of extra-articular proximal phalangeal fractures.

METHODS

Midshaft osteotomies were performed on 21 cadaveric proximal phalanges. The phalanges were fixed via dorsal plating using either unicortical or bicortical locking screws and subjected to a three-point bending test. Cyclical loading was performed, increasing displacement by 0.4 mm every block of five cycles, until construct failure. Clinical failure was defined as 2 mm of displacement. Data collected included maximal force at failure, force at clinical failure, number of cycles to failure, and displacement to failure.

RESULTS

Similar values were noted in maximal force at clinical failure and biomechanical failure, number of cycles to failure, and displacement to failure between phalanges fixed using unicortical or bicortical locking screws.

CONCLUSIONS

Both unicortical and bicortical locking screws provide sufficient stability to facilitate early motion without risk of clinically relevant displacement during plate fixation of midshaft proximal phalangeal fractures.

CLINICAL RELEVANCE: Unicortical dorsal plate-screw constructs should be considered, where appropriate, to prevent tendon injury or complications.

Для цитирования: Sivakumar Brahman Shankar, Bissell Lianne, Hile Mark, Clarke Elizabeth. Unicortical Versus Bicortical Screws for Dorsal Plate Fixation of Phalangeal Fractures.. The Journal of hand surgery. 2025. DOI: 10.1016/j.jhsa.2025.03.009.