DOI 10.1016/j.hrthm.2025.06.050

Медицинская библиотека · Heart rhythm · 2025

Human use conditions in fatigue of intravenous leads used in cardiac implanted electronic devices.

Quinn Timothy, Swerdlow Charles D, Koepke Lucas, Dawson James, Barka Nicki Mara, Santelli Jason, Muff Diane, Savela Jordan, Felber Alex, Friedrich Michael, Rump Jens, Luther Torsten, Benzing Jake, Peterson Michael, Chawla Mohit, Weiner Stanislav, Colley B Judson, Piccini Jonathan P

СтатьяМногоцентровое исследованиеНаблюдательное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Transvenous lead conductors are subject to fatigue fracture from repetitive bending stresses, which are proportional to alternating curvature. A prerequisite for developing a valid preclinical lead-testing standard is knowledge of the stresses on implanted leads - their fatigue environment, but little is known about it.

OBJECTIVE

The Human Use Condition Study (HUCS) quantified this fatigue environment, including the relationship between lead stiffness and alternating curvature.

METHODS

HUCS was a prospective, observational, multicenter clinical study. Leads from 4 manufacturers were chosen to span the range of stiffness in clinical use. They were imaged using biplane cinefluoroscopy during cardiac motion and in the shoulder region during arm motion. In each image, the lead was traced, and its curvature was calculated as a function of time.

RESULTS

Images were analyzed for 109 subjects. In each region of the lead analyzed, maximum alternating curvature was determined primarily by between-patient differences, not differences in lead stiffness. Across regions, maximum alternating curvature was greatest in the extravenous region (P = .001). In this region, the length of the highly stressed portion of the lead correlated inversely with lead stiffness (P < .001).

CONCLUSION

HUCS is the first study to measure alternating in vivo stresses applied to leads from multiple manufacturers that spanned a range of stiffnesses and thereby enable the development of an evidence-based lead-testing standard. The greatest alternating stress occurs in the extravenous region. There, lead stiffness predicts how much of the lead is highly stressed but does not predict maximum alternating curvature.

Для цитирования: Quinn Timothy, Swerdlow Charles D, Koepke Lucas, Dawson James, Barka Nicki Mara, Santelli Jason, Muff Diane, Savela Jordan, Felber Alex, Friedrich Michael, Rump Jens, Luther Torsten, Benzing Jake, Peterson Michael, Chawla Mohit, Weiner Stanislav, Colley B Judson, Piccini Jonathan P. Human use conditions in fatigue of intravenous leads used in cardiac implanted electronic devices.. Heart rhythm. 2025. DOI: 10.1016/j.hrthm.2025.06.050.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.