MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00264-016-3363-4

Медицинская библиотека · International orthopaedics · 2016

Dual mobility cups in revision total hip arthroplasty.

Viste Anthony, Desmarchelier Romain, Fessy Michel-Henri

СтатьяПеревод в очереди

Аннотация

AIM: The purpose of our study was to evaluate the dislocation and aseptic loosening rates of a dual mobility construct in revision total hip arthroplasty (THA).

METHODS

Three hundred thirty-four revision THAs performed between 2006 and 2011, using a dual mobility cup, were included in this retrospective study. The indications for revision were aseptic loosening (70%), infection (21%) and recurrent instability (9%). The minimum follow-up was five years (mean 7 ± 2 years, maximum 10 years).

RESULTS

At the latest follow-up, 11 episodes of dislocation occurred, seven of which were recurrent (1.3% for aseptic loosening, 2.8% for infection and 3.4% for recurrent dislocation). Ten cases of aseptic loosening occurred; in seven of them, a cemented cup into a reinforcement ring had been used (OR = 14, p = 0.0001).

CONCLUSION

This study provided evidence of the advantages of dual mobility cups in all revision THA indications.

Для цитирования: Viste Anthony, Desmarchelier Romain, Fessy Michel-Henri. Dual mobility cups in revision total hip arthroplasty.. International orthopaedics. 2016. DOI: 10.1007/s00264-016-3363-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление