Latarjet procedure: biomechanical evaluation of 2-screw coracoid fixation.
Аннотация
BACKGROUND
Coracoid nonunion is a relevant complication following the Latarjet procedure and is influenced by multiple factors, including the method of graft fixation. The purpose of this study was to evaluate and characterize the biomechanical properties of various two-screw fixation constructs used for coracoid graft fixation in the Latarjet procedure.
METHODS
Forty model scapulae (Sawbones Inc., Vashon, WA, USA) were used for this study. A 15% anterior inferior glenoid bone defect was created. The coracoid was osteotomized at the juncture of the vertical and horizontal aspects, transferred to the anterior-inferior edge of the glenoid, and fixed with either two 3.5 mm fully threaded cannulated cortical screws, two 3.5 mm fully threaded solid cortical screws, two 3.5 mm partially threaded cannulated screws, or two 4.5 mm partially threaded malleolar screws (MS). Biomechanical testing was performed with an Instron material testing machine (Instron Corp., Norwood, MA, USA) by applying loads to the lateral aspect of the transferred coracoid graft. The constructs were preconditioned with nondestructive cyclical loading (0N-20N) to determine construct stiffness. After 100 cycles of dynamic loading, the construct was loaded to failure to determine ultimate failure load, yield displacement, and mode of failure.
RESULTS
All failures were associated with plastic deformation of the screws and coracoid graft fracture. There was a significantly lower initial stiffness for partially threaded cannulated screws compared to MS (186 ± 49.3 N/mm vs. 280 ± 65.5 N/mm, P = .01) but no significant differences among the other constructs. There was no difference in ultimate failure load (P = .18) or yield displacement (P = .05) among constructs.
CONCLUSION
Two screw coracoid fixation of the coracoid in a simulated classic Latarjet procedure with 3.5 mm fully threaded cortical and cannulated screws is comparable to 4.5 mm MS in strength, stiffness, and displacement at failure. On the other hand, partially threaded 3.5 mm cannulated screws provide inferior fixation stiffness and could potentially affect clinical outcomes.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.