Медицинская библиотека · European journal of trauma and emergency surgery : official publication of the European Trauma Society · 2007
Secondary Collapse of an Expandable Cage After Vertebral Corpectomy.
СтатьяПеревод в очереди
Аннотация
Expandable vertebral body replacement systems have been increasingly used for anterior stabilization of spine. We report a secondary collapse of an expandable vertebral body replacement system. This specific complication has not been reported in the literature so far. The most obvious reason for failure was insufficient tightening of a locking screw. This paper emphasizes the importance of correct technical application.
Для цитирования: Sattler Martin, Goesling Thomas, Busche Marc, Krettek Christian, Bastian Leonard. Secondary Collapse of an Expandable Cage After Vertebral Corpectomy.. European journal of trauma and emergency surgery : official publication of the European Trauma Society. 2007. DOI: 10.1007/s00068-007-6021-0.