DOI 10.2106/jbjs.20.00833

Медицинская библиотека · The Journal of bone and joint surgery. American volume · 2021

Relationship Between Glenoid Component Shift and Osteolysis After Anatomic Total Shoulder Arthroplasty: Three-Dimensional Computed Tomography Analysis.

Ricchetti Eric T, Jun Bong-Jae, Jin Yuxuan, Ho Jason C, Patterson Thomas E, Dalton Jarrod E, Derwin Kathleen A, Iannotti Joseph P

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The purpose of this study was to evaluate glenoid component position and radiolucency following anatomic total shoulder arthroplasty (TSA) using sequential 3-dimensional computed tomography (3D CT) analysis.

METHODS

In a series of 152 patients (42 Walch A1, 16 A2, 7 B1, 49 B2, 29 B3, 3 C1, 3 C2, and 3 D glenoids) undergoing anatomic TSA with a polyethylene glenoid component, sequential 3D CT analysis was performed preoperatively (CT1), early postoperatively (CT2), and at a minimum 2-year follow-up (CT3). Glenoid component shift was defined as a change in component version or inclination of ≥3° from CT2 to CT3. Glenoid component central anchor peg osteolysis (CPO) was assessed at CT3. Factors associated with glenoid component shift and CPO were evaluated.

RESULTS

Glenoid component shift occurred from CT2 to CT3 in 78 (51%) of the 152 patients. CPO was seen at CT3 in 19 (13%) of the 152 patients, including 15 (19%) of the 78 with component shift. Walch B2 glenoids with a standard component and glenoids with higher preoperative retroversion were associated with a higher rate of shift, but not of CPO. B3 glenoids with an augmented component and glenoids with greater preoperative joint-line medialization were associated with CPO, but not with shift. More glenoid component joint-line medialization from CT2 to CT3 was associated with higher rates of shift and CPO. A greater absolute change in glenoid component inclination from CT2 to CT3 and a combined absolute glenoid component version and inclination change from CT2 to CT3 were associated with CPO. Neither glenoid component shift nor CPO was associated with worse clinical outcomes.

CONCLUSIONS

Postoperative 3D CT analysis demonstrated that glenoid component shift commonly occurs following anatomic TSA, with increased inclination the most common direction. Most (81%) of the patients with glenoid component shift did not develop CPO. Longer follow-up is needed to determine the relationships of glenoid component shift and CPO with loosening over time.

LEVEL OF EVIDENCE: Therapeutic Level IV. See Instructions for Authors for a complete description of levels of evidence.

Для цитирования: Ricchetti Eric T, Jun Bong-Jae, Jin Yuxuan, Ho Jason C, Patterson Thomas E, Dalton Jarrod E, Derwin Kathleen A, Iannotti Joseph P. Relationship Between Glenoid Component Shift and Osteolysis After Anatomic Total Shoulder Arthroplasty: Three-Dimensional Computed Tomography Analysis.. The Journal of bone and joint surgery. American volume. 2021. DOI: 10.2106/jbjs.20.00833.

Предметные рубрики MeSH