Biomechanical analysis of orthogonal and unilateral locking plate constructs in a fracture gap model.

Welsh Peter J, Nylund Adam M, Gilbert Peter J, Smith Nick A, Smith Lloyd V

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To compare unilaterally plated (UP) constructs and orthogonally plated (OP) constructs of differing sizes (3.5 mm + 2.0 mm locking plates [OP2.0], 3.5 mm + 2.4 mm locking plates [OP2.4] and 3.5 mm + 3.0 mm locking plates [OP3.0]) under cyclic and static axial compression load to failure testing using a fracture gap model.

STUDY

DESIGN

In vitro experimental biomechanical study.

SAMPLE POPULATION: Acetal homopolymer (Delrin) rods stabilized with locking plates and screws (Arthrex OrthoLine).

METHODS

One of four stabilization techniques (UP, OP2.0, OP2.4, OP3.0) was applied to rods with a fixed fracture gap. Constructs were fatigue tested under axial compression (90,000 cycles; 4-196 N) followed by static load to failure. Cyclic displacement was evaluated after the first, middle, and last 100 cycles. Stiffness and strength were analyzed during static axial compression load to failure.

RESULTS

During cyclic testing, UP experienced 3.5, 3.8, and 4.1 times the gap strain of OP2.0, OP2.4, and OP3.0, respectively (p < 0.0075). Fatigue and construct design had significant effects on displacement (p < .0001). OP2.0, OP2.4, and OP3.0 demonstrated 2.5, 3.0, and 4.1 times the strength and 3.0, 3.6, and 4.2 times the stiffness of UP, respectively (p < .0002).

CONCLUSION

In the present in vitro fracture gap model, OP constructs were stronger and stiffer than UP under dynamic and static axial compression, and OP stiffness increased with increasing implant size.

CLINICAL SIGNIFICANCE: The results of this study demonstrate objective biomechanical advantages of OP compared to UP. Based on these results, orthogonal plating can be considered when increased fixation strength and stiffness are warranted.

Для цитирования: Welsh Peter J, Nylund Adam M, Gilbert Peter J, Smith Nick A, Smith Lloyd V. Biomechanical analysis of orthogonal and unilateral locking plate constructs in a fracture gap model.. Veterinary surgery : VS. 2025. DOI: 10.1111/vsu.14206.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.