MedElevator Войти в исследования
← Поиск публикаций

DOI 10.17245/jdapm.2023.23.5.281

Медицинская библиотека · Journal of dental anesthesia and pain medicine · 2023

Bending 30-gauge needles using a needle guide: fatigue life evaluation.

Tuttle Jared Joseph, Davidson Andrew Doran, Tuttle Gregory Kent

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Dentists bend needles prior to certain injections; however, there are concerns regarding needle fracture, lumen occlusion, and sharps handling. A previous study found that a 30-gauge needle fractures after four to nine 90° bends. This fatigue life study evaluated how many 90° bends a 30-gauge dental needle will sustain before fracture when bent using a needle guide.

METHODS

Two operators at Element Materials Technology, an independent testing, inspection, and certification company tested 48 30-gauge needles. After applying the needle guide, the operators bent the needle to a 90° angle and expressed the anesthetic from the tip. The needle was then bent back to a 0° angle, and the functionality was tested again. This process was repeated until the anesthetic failed to pass through the end of the needle due to fracture or obstruction. Each operator tested 24 needles (12 needles from each lot), and the number of sustained bends before the needle fracture was recorded.

RESULTS

The average number of sustained bends before needle failure was 40.33 (95% confidence interval = 37.41-43.26), with a minimum of 20, median of 40, and a maximum of 54. In each trial, the lumen remained patent until the needle fractured. The difference between the operators was statistically significant (P < 0.001). No significant differences in performance between needle lots were observed (P = 0.504).

CONCLUSION

Our results suggest that using a needle guide increases the number of sustained bends before needle fracture (P < 0.000001) than those reported in previous studies. Future studies should further evaluate the use of needle guides with other needle types across a variety of operators. Furthermore, additional opportunities lie in exploring workplace safety considerations and clinical applications of anesthetic delivery using a bent needle.

Для цитирования: Tuttle Jared Joseph, Davidson Andrew Doran, Tuttle Gregory Kent. Bending 30-gauge needles using a needle guide: fatigue life evaluation.. Journal of dental anesthesia and pain medicine. 2023. DOI: 10.17245/jdapm.2023.23.5.281.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.