Multiplanar Semicircular New-Generation Implant System Developed for Proximal Femur Periprosthetic Fractures: A Biomechanical Study.

Satılmış Ahmet Burak, Ülker Ahmet, Uzunay Zafer, Cengiz Tolgahan, Temiz Abdurrahim, Yaşar Mustafa, Mutlu Tansel, Daşar Uygar

СтатьяПеревод в очереди

Аннотация

Background and

Objectives

The study aimed to evaluate a newly designed semicircular implant for the fixation of Vancouver Type B1 periprosthetic femoral fractures (PFFs) in total hip arthroplasty (THA) patients. To determine its strength and clinical applicability, the new implant was compared biomechanically with conventional fixation methods, such as lateral locking plate fixation and a plate combined with cerclage wires. Materials and

Methods

Fifteen synthetic femur models were used in this biomechanical study. A Vancouver Type B1 periprosthetic fracture was simulated by osteotomy 5 mm distal to the femoral stem. The models were divided into three groups: Group I (lateral locking plate fixation), Group II (lateral locking plate with cerclage wires), and Group III (new semicircular implant system). All fixation methods were subjected to axial loading, lateral bending, and torsional force testing using an MTS biomechanical testing device. Failure load and displacement were measured to assess stability.

Results

The semicircular implant (Group III) demonstrated a significantly higher failure load (778.8 ± 74.089 N) compared to the lateral plate (Group I: 467 ± 68.165 N) and plate with cerclage wires (Group II: 652.4 ± 65.474 N; p < 0.001). The new implant also exhibited superior stability under axial, lateral bending, and torsional forces. The failure load for Group III was more robust, with fractures occurring at the screw level rather than plate or screw detachment.

Conclusions

Compared to traditional fixation methods, the newly designed semicircular implant demonstrated superior biomechanical performance in stabilizing Vancouver Type B1 periprosthetic femoral fractures. It withstood higher physiological loads, offered better structural stability, and could be an alternative to existing fixation systems in clinical practice. Further studies, including cadaveric and in vivo trials, are recommended to confirm these results and assess the long-term clinical outcomes.

Для цитирования: Satılmış Ahmet Burak, Ülker Ahmet, Uzunay Zafer, Cengiz Tolgahan, Temiz Abdurrahim, Yaşar Mustafa, Mutlu Tansel, Daşar Uygar. Multiplanar Semicircular New-Generation Implant System Developed for Proximal Femur Periprosthetic Fractures: A Biomechanical Study.. Medicina (Kaunas, Lithuania). 2025. DOI: 10.3390/medicina61010110.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.