MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1186/s13018-020-01926-w

Медицинская библиотека · Journal of orthopaedic surgery and research · 2020

Does new instrument for Oxford unicompartmental knee arthroplasty improve short-term clinical outcome and component alignment? A meta-analysis.

Sun Xiao Wei, Lu Fei Fan, Zou Kun, Hong Mao, Zhang Qi Dong, Guo Wan Shou

СтатьяМетаанализОбзорПеревод в очереди

Аннотация

BACKGROUND

The Microplasty (MP) instrumentation designed for the Phase III Oxford mobile-bearing unicompartmental knee arthroplasty (UKA) system is considered a better option to achieve more accurate component positioning and alignment. In the present study, we focused on short-term clinical and radiological outcomes to determine whether the MP instrumentation can reduce the short-term revision rate and occurrence of outliers of metallic components.

METHODS

The literature in PubMed, Embase, the Cochrane Library, and Web of Science was searched up to May 2020. Studies were scrutinized by two independent authors, and the revision rate, complication spectrum, and radiological assessment with outlier rates were specifically analyzed. RevMan 5.3 was used for the statistical analysis.

RESULTS

Seven studies were included in the meta-analysis. Four studies reported both clinical and radiological outcomes, two reported only radiological outcomes, and one reported only clinical outcomes. The pooled analysis showed that the revision rate in the MP instrumentation group was 0.866 per 100 component years, while that in the control group was 1.124 (odds ratio, 0.77; p < 0.05). The subgroup analysis of the bearing dislocation rate showed a significantly greater reduction in the Korean population than in the populations of other countries (p < 0.05). The radiological assessment showed that the alignment of the femoral component was significantly improved (p < 0.05), while that of the tibial component was not (p > 0.05).

CONCLUSION

The newly developed MP instrumentation for Oxford UKA significantly reduced the revision rate of this treatment. The positioning of the femoral component was also proven to be better by radiological assessments.

Для цитирования: Sun Xiao Wei, Lu Fei Fan, Zou Kun, Hong Mao, Zhang Qi Dong, Guo Wan Shou. Does new instrument for Oxford unicompartmental knee arthroplasty improve short-term clinical outcome and component alignment? A meta-analysis.. Journal of orthopaedic surgery and research. 2020. DOI: 10.1186/s13018-020-01926-w.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.