DOI 10.1016/j.injury.2024.111512
Comparative biomechanical analysis of reconstruction and cephalomedullary nails in the treatment of osteoporotic subtrochanteric fractures.
Аннотация
INTRODUCTION
This study aimed to compare the biomechanical properties of two types of intramedullary nails - reconstruction nails (RCN) and cephalomedullary nails (CMN) - each with different proximal fixations, in a model of an osteoporotic subtrochanteric femoral fracture. This study focused on assessing stiffness and load to failure of RCN and CMN nails to provide insight into their clinical applications in osteoporotic fracture treatments.
MATERIALS AND
METHODS
Ten synthetic osteoporotic femoral models were used to generate a comminuted subtrochanteric fracture model. Five femurs were fixed using an RCN, and the remaining five were fixed using a CMN. The constructs were subjected to axial compression to measure their structural stiffness, load to failure, and failure modes.
RESULTS
The CMN group demonstrated a slightly higher load to failure (mean, 2250 N) than the RCN group (mean, 2100 N), which was statistically significant (p = 0.008). However, the stiffness in both groups was statistically similar (RCN, 250 N/mm; CMN, 255 N/mm; p = 0.69). Both groups showed a load to failure exceeding 1500 N, a typically exerted load on the femoral head by a 75 kg individual. The failure patterns differed, with CMN failures starting at the nail insertion area and RCN failures starting at the reconstruction screw area.
CONCLUSION
The RCN offers stiffness comparable to that of the CMN; although its load to failure is slightly lower than that of the CMN, it still exceeds the physiological tolerance limit. These findings suggest that the RCN is a viable alternative for treating osteoporotic subtrochanteric fractures.