DOI 10.1016/j.echo.2009.04.001
Acute failure of a St. Jude's prosthetic aortic valve: large pannus formation masked by a small thrombus.
Аннотация
Pannus formation and valve thrombus can cause prosthetic valve failure. The authors report the case of a 50-year-old woman who presented to the emergency room with decompensated heart failure secondary to mechanical valve dysfunction. On two-dimensional and transesophageal echocardiography, the patient had severe aortic stenosis and regurgitation. A thrombus seen on the valve was felt to be the etiology of her prosthetic valve failure. She underwent emergent cardiac surgery for aortic valve replacement. Pathology revealed that although a small thrombus was present, extensive pannus was the underlying mechanism of valve dysfunction. Differentiation between pannus and thrombus may have important clinical implications, but this case illustrates that distinguishing between these entities by echocardiographic and clinical criteria may not be possible.