DOI 10.1097/00007632-199808150-00016
Failure of a carbon fiber implant. A case report.
Аннотация
STUDY
DESIGN
Case report.
OBJECTIVES
Failure of a carbon fiber implant.
SUMMARY OF BACKGROUND DATA: To simplify the procedure of posterior lumbar interbody fusion, a carbon-fiber-reinforced polymer implant has been developed. The implant has ridges to resist retropulsion, struts to support weight, and a hollow area to allow packing of autologous bone graft. So far, no complications have been reported from the use of carbon implant as a fusion aid in spine surgery.
METHODS
A patient with postoperative infection has been followed with computed tomography images and histologic examination from a reoperation.
RESULTS
An entire nonunion across the width of the disc space and a clearly broken cage was visualized with computed tomography. The spinal canal was explored during a reoperation and the tissue surrounding the dura and nerves were all black. Microscopic examination showed a large quantity of carbon particulate debris. The authors have operated on approximately 100 patients so far and no other carbon cage has broken, to their knowledge.
CONCLUSIONS
Carbon cages can break if a nonunion occurs and as a result free carbon particles move out to the spinal canal.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.