DOI 10.1016/j.arth.2020.05.065
Similar Risk of Revision After Kinematically Aligned, Patient-Specific Instrumented Total Knee Arthroplasty, and All Other Total Knee Arthroplasty: Combined Results From the Australian and New Zealand Joint Replacement Registries.
Аннотация
BACKGROUND
Kinematic alignment (KA) in total knee arthroplasty (TKA) replicates individual knee joint line orientation and physiological soft tissue balance. There are limited data on the effect of KA on survivorship compared with other methods of implanting a TKA. The aim of this study was to compare the survivorship of the same design of TKA implanted with either patient-specific instrument (PSI) KA or with computer-assisted surgery and conventionally instrumented.
METHODS
We performed an observational study of data from 2 national joint registries on all patients who underwent a TKA with a Triathlon cruciate-retaining TKA (Stryker, Mahwah, NJ) from January 2011 to December 31, 2013. The outcomes of knees performed with unrestricted KA using patient-specific instrument were compared with computer-assisted surgery and conventionally instrumented Triathlon cruciate-retaining TKA. The principal outcome measure was time to first revision using Kaplan-Meier estimates of survivorship. Cox proportional hazard ratios, adjusted for age and gender, were used to compare the 2 groups.
RESULTS
There were 20,512 TKA procedures recorded, of which 416 were performed using KA-PSI. The KA-PSI group was younger (66.8 ± 8.2 vs 68.3 ± 9.2, P = .001), with an identical gender distribution. The cumulative revision rate at 7 years was 3.1% for the KA-PSI Triathlon cohort and 3.0% for the other Triathlon TKA cohort (P = .89). The most common reasons for revision in the KA-PSI group were maltracking, patella erosion, and arthrofibrosis.
CONCLUSION
Kinematically aligned Triathlon TKA using PSI has similar revision rate as all other Triathlon TKA.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.