BACKGROUND

It is unclear whether a connection exists between femoral head size, offset, neck length, and cup abduction angles, and rate of revision in metal-on-metal (MoM) total hip arthroplasty (THA) implant systems.

METHODS

A retrospective review of MoM THA completed by a single surgeon with a single implant between 2003 and 2008 was conducted. Patient demographics, implant data, radiographs, and revision details were collected at follow-up. Incidence rates for revision and osteolysis were calculated in regard to the femoral head size, stem offset, neck length, and cup abduction angles.

RESULTS

Six hundred and ninety two THAs were identified, with 79% of patients returning for a median follow-up of 10.3 years (interquartile range = 6.0-12.3). The median time to revision was 7.5 years (interquartile range = 5.3-9.9) among 27 total revision surgeries. The overall incidence rate of revision was 5.4 revisions per 1000 person-years, 3.0 revisions per 1000 person-years for adverse local tissue reaction. Hips with a cup abduction angle of ≤40° had revisions at nearly twice the rate of those with an angle of 41°-50° (incidence rate ratio = 1.98, 95% confidence interval: 0.92, 4.29). Hips with a 9 mm neck length had an increased rate of revision (incidence rate ratio = 5.94, 95% confidence interval: 1.33, 26.55) relative to those with a neck length of 0 mm. Rates of osteolysis were similar between implants of different head sizes, neck lengths and cup abduction angles.

CONCLUSIONS

MoM implant systems with longer necks and smaller cup abduction angles may lead to increased need for revision. Results from this study suggest a need for closer long-term follow-up of MoM THA systems.

Для цитирования: LaHaise Kristen, Mandell Daniel, Golenbock Samuel, Hopkins Kassandra, Mattingly David. Pinnacle Ultamet metal-on-metal total hip arthroplasty survivorship: average 10-year follow-up.. Arthroplasty today. 2020. DOI: 10.1016/j.artd.2020.02.010.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.