Аннотация

BACKGROUND

In the setting of periprosthetic humeral fractures, the humeral stem of the implant represents a substantial challenge to the optimal method of proximal fixation. This study aimed to compare the initial biomechanical stability provided by cerclage cables with a locking plate insert versus bicortical locking screws (i.e., the gold standard for fixation) in fresh cadaveric humeri.

METHODS

After calculating the sample size, we utilized 10 sets of cadaveric specimens and created a 5-mm osteotomy gap 120 mm distal to the tip of the greater tuberosity, simulating a Wright and Cofield type-B periprosthetic humeral fracture on each specimen. Using 3 locking screws for distal fragment fixation, identical in all specimens, the specimens were assigned to Group A (3 cerclage cables with a plate insert) or Group B (3 locking bicortical screws) for proximal fragment fixation. Biomechanical tests included stiffness in varus and valgus bending, torsion, and axial compression, and a single load to failure.

RESULTS

No significant differences were observed in the biomechanical metrics between the 2 groups.

CONCLUSIONS

Our study revealed that fixation with use of cerclage cables with a plate insert demonstrated biomechanical stability comparable with that of bicortical locking screw fixation when addressing the proximal fragment in Wright and Cofield type-B periprosthetic humeral fractures.

CLINICAL RELEVANCE: For proximal fragment fixation of periprosthetic humeral fractures, cerclage cables with a plate insert can be utilized as an effective fixation method that offers initial fixation strength that is comparable to the use of 3 locking bicortical screws.

Для цитирования: Lee Yong-Jun, Lim Joon-Ryul, Choi Jee-Hoon, Yoon Tae-Hwan, Choi Yun-Rak, Chun Yong-Min. Comparing the Biomechanical Stability of Cerclage Cable with Plate Insert Versus Locking Screw in Periprosthetic Humeral Fracture.. The Journal of bone and joint surgery. American volume. 2024. DOI: 10.2106/jbjs.23.00830.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.