Biomechanical compression tests with a new implant for thoracolumbar vertebral body replacement.

Knop C, Lange U, Bastian L, Oeser M, Blauth M

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

The authors present an investigation into the biomechanical functioning of a new titanium implant for vertebral body replacement (Synex). Possible indications are fractures and/or dislocations with damage of the anterior column, posttraumatic kyphosis and tumors of the thoracolumbar spine. The construction must be supplemented by a stabilizing posterior or anterior implant. For best fit and contact with adjacent end-plates, Synex is distractable in situ. We performed comparative compression tests with Synex and MOSS ("Harms mesh cage") on human cadaveric specimens of intact vertebrae (L1). The aim of the study was to measure the compressive strength of the vertebral body end-plate in uniaxial loading via both implants to exclude collapse of Synex in vivo. Twelve human cadaveric specimens of intact vertebrae (L1) were divided into two identical groups (matched pairs) according to bone mineral density (BMD), determined using dual-energy quantitative computed tomography (DE-QCT). The specimens were loaded with an axial compression force at a constant speed of 5 mm/min to failure, and the displacement was recorded with a continuous load-displacement curve. The mean ultimate compression force (Fmax) showed a tendency towards a higher reading for Synex: 3396 N versus 2719 N (non-significant). The displacement until Fmax was 2.9 mm in the Synex group, which was half as far as in the MOSS group (5.8 mm). The difference was significant (P < 0.001). The compression force was twice as high, and significantly (P < 0.05) higher with Synex at displacements of 1 mm, 1.5 mm and 2 mm. A significant (P < 0.001) correlation (R = 0.89) between Fmax and BMD was found. Synex was found to be at least comparable to MOSS concerning the compressive performance at the vertebral end-plate. A possible consequence of the significantly higher mean compression forces between 1 and 2 mm displacement might be decreased collapse of the implant into the vertebral body in vivo.

Для цитирования: Knop C, Lange U, Bastian L, Oeser M, Blauth M. Biomechanical compression tests with a new implant for thoracolumbar vertebral body replacement.. European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society. 2001. DOI: 10.1007/s005860000211.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.