Anterior Lumbar Interbody Fusion Integrated Screw Cages: Intrinsic Load Generation, Subsidence, and Torsional Stability.
Аннотация
OBJECTIVE
To perform a repeatable idealized in vitro model to evaluate the effects of key design features and integrated screw fixation on unloaded surface engagement, subsidence, and torsional stability.
METHODS
We evaluated four different stand-alone anterior lumbar interbody fusion (ALIF) cages with two, three, and four screw designs. Polyurethane (saw-bone) foam blocks were used to simulate the vertebral bone. Fuji Film was used to measure the contact footprint, average pressure, and load generated by fixating the cages with screws. Subsidence was tested by axially loading the constructs at 10 N/s to 400 N and torsional load was applied +/-1 Nm for 10 cycles to assess stability. Outcome measures included total subsidence and maximal torsional angle range.
RESULTS
Cages 1, 2, and 4 were symmetrical and produced similar results in terms of contact footprint, average pressure, and load. The addition of integrated screws into the cage-bone block construct demonstrated a clear trend towards decreased subsidence. Cage 2 with surface titanium angled ridges and a keel produced the greatest subsidence with and without screws, significantly more than all other cages ( P < 0.05). Angular rotation was not significantly affected by the addition of screws ( P < 0.066). A statistically significant correlation existed between subsidence and reduced angular rotation across all cage constructs ( P = 0.018).
CONCLUSION
Each stand-alone cage featured unique surface characteristics, which resulted in differing cage-foam interface engagement, influencing the subsidence and torsional angle. Increased subsidence significantly reduced the torsional angle across all cage constructs.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.