DOI 10.1097/brs.0000000000001985

Медицинская библиотека · Spine · 2017

Is It "In" or "Out"? The Optimal Fluoroscopic Views for Intraoperative Determination of Proper Lateral Mass Screw Placement.

Kim Sang Bum, Rhee John M, Oh Byung Hak, Won You Gun, Jung Yousun, Park Kun Young, Hutton William C, Kim Chulmin

СтатьяПеревод в очереди

Аннотация

STUDY

DESIGN

Cadaveric.

OBJECTIVE

Determine optimal fluoroscopic views for detecting cervical lateral mass screw (LMS) violations.

SUMMARY OF BACKGROUND DATA: Single plane intraoperative x-rays are commonly used but frequently inadequate due to its complex trajectory. Fluoroscopy can be taken in multiple planes, but the ideal fluoroscopic view to assess malposition is not known: depending on the view, any given screw may look "in" or "out."

METHODS

C3-6 LMS were inserted in three cadavers. To evaluate neuroforaminal violation, LMS were inserted into the foramen with the tip penetrating the anterior cortex by 0, 2, and 4 mm. To assess facet joint violation, LMS were inserted toward the subjacent facet joint with the tip penetrating the anterior cortex by 0 and 2 mm. Fluoroscopic views were taken 0°, 10°, 20°, 30°, and 40° to the lateral plane. Views were independently evaluated by three blinded spine surgeons.

RESULTS

Twenty-degree oblique view correctly identified a 2 mm penetration into the neuroforamen in 79%, and a 4 mm penetration in 86%, for a sensitivity of 83% and specificity of 90%. Thirty-degree view had lower sensitivity (76%) but slightly higher specificity (93%). Twenty-degree and 30° views were significantly more sensitive than the other views. Zero-degree view correctly identified a 2 mm penetration into the facet joint in 93%, for a sensitivity of 93% and specificity of 92%. Ten-degree view had lower sensitivity (72%) but higher specificity (100%). The 0° view was significantly more sensitive than the other views.

CONCLUSION

Twenty-degree and 30° oblique views significantly provided the most sensitive assessment of LMS potentially violating the neuroforamen, whereas the 0° neutral lateral view significantly provided the most sensitive assessment of facet violations. The specificities were also high (in the 90% range) for these views. We recommend the use of these views intraoperatively when assessing proper placement of LMS fluoroscopically.

LEVEL OF EVIDENCE: N/A.

Для цитирования: Kim Sang Bum, Rhee John M, Oh Byung Hak, Won You Gun, Jung Yousun, Park Kun Young, Hutton William C, Kim Chulmin. Is It "In" or "Out"? The Optimal Fluoroscopic Views for Intraoperative Determination of Proper Lateral Mass Screw Placement.. Spine. 2017. DOI: 10.1097/brs.0000000000001985.