MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1038/s41390-025-04179-7

Медицинская библиотека · Pediatric research · 2025

Non-invasive meningitis screening in neonates and infants: multicentre international study.

Ajanovic Sara, Jobst Beatrice, Jiménez Javier, Quesada Rita, Santos Fabião, Carandell Francesc, Lopez-Azorín Manuela, Valverde Eva, Ybarra Marta, Bravo María Carmen, Petrone Paula, Sial Hassan, Muñoz David, Agut Thais, Salas Barbara, Carreras Nuria, Alarcón Ana, Iriondo Martín, Luaces Carles, Sidat Muhammad, Zandamela Mastalina, Rodrigues Paula, Graça Dulce, Ngovene Sebastião, Bramugy Justina, Cossa Anelsio, Mucasse Campos, Buck William Chris, Arias Sara, El Abbass Chaymae, Tligi Houssain, Barkat Amina, Ibáñez Alberto, Parrilla Montserrat, Elvira Luis, Calvo Cristina, Pellicer Adelina, Cabañas Fernando, Bassat Quique, UNITED study group

СтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

BACKGROUND AND

OBJECTIVES

Meningitis diagnosis requires a lumbar puncture (LP) to obtain cerebrospinal fluid (CSF) for a laboratory-based analysis. In high-income settings, LPs are part of the systematic approach to screen for meningitis, and most yield negative results. In low- and middle-income settings, LPs are seldom performed, and suspected cases are often treated empirically. The aim of this study was to validate a non-invasive transfontanellar white blood cell (WBC) counter in CSF to screen for meningitis.

METHODS

We conducted a prospective study across three Spanish hospitals, one Mozambican and one Moroccan hospital (2020-2023). We included patients under 24 months with suspected meningitis, an open fontanelle, and a LP performed within 24 h from recruitment. High-resolution-ultrasound (HRUS) images of the CSF were obtained using a customized probe. A deep-learning model was trained to classify CSF patterns based on LPs WBC counts, using a 30cells/mm3 threshold.

RESULTS

The algorithm was applied to 3782 images from 76 patients. It correctly classified 17/18 CSFs with ≥ 30 WBC, and 55/58 controls (sensitivity 94.4%, specificity 94.8%). The only false negative was paired to a traumatic LP with 40 corrected WBC/mm3.

CONCLUSIONS

This non-invasive device could be an accurate tool for screening meningitis in neonates and young infants, modulating LP indications.

IMPACT: Our non-invasive, high-resolution ultrasound device achieved 94% accuracy in detecting elevated leukocyte counts in neonates and infants with suspected meningitis, compared to the gold standard (lumbar punctures and laboratory analysis). This first-in-class screening device introduces the first non-invasive method for neonatal and infant meningitis screening, potentially modulating lumbar puncture indications. This technology could substantially reduce lumbar punctures in low-suspicion cases and provides a viable alternative critically ill patients worldwide or in settings where lumbar punctures are unfeasible, especially in low-income countries).

Для цитирования: Ajanovic Sara, Jobst Beatrice, Jiménez Javier, Quesada Rita, Santos Fabião, Carandell Francesc, Lopez-Azorín Manuela, Valverde Eva, Ybarra Marta, Bravo María Carmen, Petrone Paula, Sial Hassan, Muñoz David, Agut Thais, Salas Barbara, Carreras Nuria, Alarcón Ana, Iriondo Martín, Luaces Carles, Sidat Muhammad, Zandamela Mastalina, Rodrigues Paula, Graça Dulce, Ngovene Sebastião, Bramugy Justina, Cossa Anelsio, Mucasse Campos, Buck William Chris, Arias Sara, El Abbass Chaymae, Tligi Houssain, Barkat Amina, Ibáñez Alberto, Parrilla Montserrat, Elvira Luis, Calvo Cristina, Pellicer Adelina, Cabañas Fernando, Bassat Quique, UNITED study group. Non-invasive meningitis screening in neonates and infants: multicentre international study.. Pediatric research. 2025. DOI: 10.1038/s41390-025-04179-7.