MedElevator Войти в исследования
← Поиск публикаций

DOI 10.2340/17453674.2024.42490

Медицинская библиотека · Acta orthopaedica · 2024

Angular stable plate fixation provides favorable biomechanical stability in simulated T-shaped acetabular fractures: a biomechanical study.

Lodde Moritz F, Katthagen Christoph, Klimek Matthias, Abshagen Karl, Peez Christian, Große-Allermann Arian, Raschke Michael J, Riesenbeck Oliver

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

BACKGROUND AND PURPOSE: The treatment of acetabular fractures remains technically demanding. In the case of reduced bone quality or fracture morphology reducing the amount of bone available for fixation, locking plates should provide considerable advantages. The aim of the present study was to compare conventional and locking plate fixation. It was hypothesized that locking plate fixation provides less displacement and higher construct stiffness.

METHODS

A T-shaped acetabular fracture was simulated in 16 synthetic pelvic models. The fracture was addressed with a biplanar 10-hole 2-column plate buttressing the medial acetabular wall. Optical markers were attached to the fracture sites for motion tracking. Standardization of the acetabulum loading mechanism was performed using a unipolar hemiarthroplasty. The primary outcome measure was displacement at the fracture sites. The secondary outcome measure was the construct stiffness (N/mm).

RESULTS

Fracture displacement was less in the group of angular stable implants compared with the group fixed with conventional non-locking implants. Under cyclic loading displacement was less in the group of locking plate fixation. No differences in mean initial axial stiffness were detected between locking plate fixation (407 N/mm) and conventional plating (308 N/mm, ∆ 99 N/mm, 95% confidence interval -48 to 245).

CONCLUSION

We showed that locking plate fixation buttressing the medial acetabular wall achieved less fracture displacement but showed no differences in axial stiffness compared with conventional plating.

Для цитирования: Lodde Moritz F, Katthagen Christoph, Klimek Matthias, Abshagen Karl, Peez Christian, Große-Allermann Arian, Raschke Michael J, Riesenbeck Oliver. Angular stable plate fixation provides favorable biomechanical stability in simulated T-shaped acetabular fractures: a biomechanical study.. Acta orthopaedica. 2024. DOI: 10.2340/17453674.2024.42490.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.