DOI 10.1186/s13018-019-1465-6

Медицинская библиотека · Journal of orthopaedic surgery and research · 2019

Patient-specific instrumentation improved three-dimensional accuracy in total knee arthroplasty: a comparative radiographic analysis of 1257 total knee arthroplasties.

Pauzenberger Leo, Munz Martin, Brandl Georg, Frank Julia K, Heuberer Philipp R, Laky Brenda, Schwameis Eva, Anderl Werner

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

The purpose of this study was to compare restoration of mechanical limb alignment and three-dimensional component-positioning between conventional and patient-specific instrumentation in total knee arthroplasty.

METHODS

Radiographic data of patients undergoing mobile-bearing total knee arthroplasty (n = 1257), using either conventional (n = 442) or patient-specific instrumentation (n = 812), were analyzed. To evaluate accuracy of axis restoration and 3D-component-positioning between conventional and patient-specific instrumentation, absolute deviations from the targeted neutral mechanical limb alignment and planned implant positions were determined. Measurements were performed on standardized coronal long-leg and sagittal knee radiographs. CT-scans were evaluated for accuracy of axial femoral implant rotation. Outliers were defined as deviations from the targeted neutral mechanical axis of > ± 3° or from the intraoperative component-positioning goals of > ± 2°. Deviations greater than ± 5° from set targets were considered to be severe outliers.

RESULTS

Deviations from a neutral mechanical axis (conventional instrumentation: 2.3°± 1.7° vs. patient-specific instrumentation: 1.7°± 1.2°; p < 0.001) and numbers of outliers (conventional instrumentation: 25.8% vs. patient-specific instrumentation: 10.1%; p < 0.001) were significantly lower in the patient-specific instrumentation group. Significantly lower mean deviations and less outliers were detected regarding 3D-component-positioning in the patient-specific instrumentation compared to the conventional instrumentation group (all p < 0.05).

CONCLUSIONS

Patient-specific instrumentation prevented from severe limb malalignment and component-positioning outliers (> ± 5° deviation). Use of patient-specific instrumentation proved to be superior to conventional instrumentation in achieving more accurate limb alignment and 3D-component positioning, particularly regarding femoral component rotation. Furthermore, the use of patient-specific instrumentation successfully prevented severe (> 5° deviation) outliers.

Для цитирования: Pauzenberger Leo, Munz Martin, Brandl Georg, Frank Julia K, Heuberer Philipp R, Laky Brenda, Schwameis Eva, Anderl Werner. Patient-specific instrumentation improved three-dimensional accuracy in total knee arthroplasty: a comparative radiographic analysis of 1257 total knee arthroplasties.. Journal of orthopaedic surgery and research. 2019. DOI: 10.1186/s13018-019-1465-6.

Предметные рубрики MeSH