Neck range of motion measurements using a new three-dimensional motion analysis system: validity and repeatability.

Inokuchi Haruhi, Tojima Michio, Mano Hiroshi, Ishikawa Yuki, Ogata Naoshi, Haga Nobuhiko

СтатьяПоддержка исследования: вне правительства СШАВалидационное исследованиеПеревод в очереди

Аннотация

PURPOSE: Neck movement is important for many activities of daily living (ADL). Neck disorders, such as cervical spondylosis and whiplash can limit neck movement and ADL. The cervical range of motion (CROM) device has been recently used to measure neck range of motion (ROM); however, this measurement includes trunk motion, and therefore does not represent a pure neck ROM measurement. The authors aimed to develop a new method to establish pure neck ROM measurements during flexion, extension, lateral bending, and rotation using a three-dimensional motion analysis system, VICON.

METHODS

Twelve healthy participants were recruited and neck ROMs during flexion, extension, lateral bending, and rotation were measured using VICON and the CROM device. Test-retest repeatability was assessed using interclass correlation coefficients (ICCs), standard error of measurement (SEM), and minimal detectable change (MDC). Validity between two measurements was evaluated using a determination coefficient and Pearson's correlation coefficient.

RESULTS

ICCs of neck ROM measured using VICON and the CROM device were all at substantial or almost perfect levels [VICON: ICC(1,2) = 0.786-0.962, the CROM device: ICC(1,2) = 0.736-0.950]. Both SEMs and MDCs were low in all measurement directions (VICON: SEM = 1.3°-4.5°, MDC = 3.6°-12.5°; the CROM device: SEM = 2.2°-3.9°, MDC = 6.1°-10.7°). Determination coefficients (R(2)s) and Pearson's correlation coefficients (rs) between the two measurement methods were high (R(2) = 0.607-0.745, r = 0.779-0.863).

CONCLUSIONS

VICON is a useful system to measure neck ROMs and evaluate the efficacy of interventions, such as surgery or physiotherapeutic exercise.

Для цитирования: Inokuchi Haruhi, Tojima Michio, Mano Hiroshi, Ishikawa Yuki, Ogata Naoshi, Haga Nobuhiko. Neck range of motion measurements using a new three-dimensional motion analysis system: validity and repeatability.. European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society. 2015. DOI: 10.1007/s00586-015-3913-2.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.