MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1186/s12891-024-07977-y

Медицинская библиотека · BMC musculoskeletal disorders · 2024

Does minimally invasive external fixation of proximal humerus fractures provide adequate stability? A biomechanical in vitro study.

Greinwald Markus, Matschl Franz, Regenbogen Stephan, Pätzold Robert, Blonna Davide, Augat Peter

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

INTRODUCTION

Proximal humerus fractures are usually treated with rigid angle-stable plates or nails. As an alternative, semi-rigid, minimally invasive systems have been developed for supportive fixation. A new system uses this principle with 2.5 mm partially threaded pins and external fixation.

AIM: This study aimed to compare the fracture stability of the fixator with an angular stable plate about whether the mechanical stability is sufficient for early postoperative mobilisation.

MATERIALS AND

METHODS

Human proximal humeri (paired, female, age ≥ 60, n = 8) were cut to a 3-part fracture and treated with an external fixator (Galaxy Shoulder, Orthofix) or an angular stable plate (Philos, DePuy Synthes). A physiological load was simulated in vitro based on a 45° abduction motion. The load level was continuously increased from 10 to 320 N within 10,000 cycles simulating the loads during rehabilitation. This was followed by another 10,000 cycles at 320 N for fatigue testing. Fracture motion was recorded with a 3D camera. Fracture stability was evaluated at a partial load of 160 N and 240 N (simulating loads during early mobilisation).

RESULTS

The fracture motion at the lateral aspect was significantly (p < 0.034) larger for the fixator compared to the plate both for axial and for shear movements, while at the medial aspects fracture motions were similar. Both fixation systems adequately stabilized the lateral tubercle fragment with no significant differences between fixation techniques. The migration of the fixator pins (1.1 ± 1.0 mm) and valgus rotation of the head fragment were minimal (3.1 ± 2.0°). The maximum load level of 320 N was reached by all plate constructs and 3 fixator constructs.

CONCLUSION

Our results suggest that the external fixator system allows early but gentle postoperative mobilisation. More aggressive rehabilitation measures should be postponed after the healing of the fracture.

Для цитирования: Greinwald Markus, Matschl Franz, Regenbogen Stephan, Pätzold Robert, Blonna Davide, Augat Peter. Does minimally invasive external fixation of proximal humerus fractures provide adequate stability? A biomechanical in vitro study.. BMC musculoskeletal disorders. 2024. DOI: 10.1186/s12891-024-07977-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.