DOI 10.1097/brs.0000000000001200

Медицинская библиотека · Spine · 2016

Pedicle Screw Fixation Under Nonaxial Loads: A Cadaveric Study.

Bianco Rohan-Jean, Aubin Carl-Eric, Mac-Thiong Jean-Marc, Wagnac Eric, Arnoux Pierre-Jean

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

STUDY

DESIGN

An experimental study of pedicle screw fixation in human cadaveric vertebrae.

OBJECTIVE

The aim of this study was to experimentally characterize pedicle screw fixation under nonaxial loading and to analyze the effect of the surgeons' screw and placement choices on the fixation risk of failure.

SUMMARY OF BACKGROUND DATA: Pedicle screw fixation performance is traditionally characterized with axial pullout tests, which do not fully represent the various tridimensional loads sustained by the screws during correction maneuvers of severe spinal deformities. Previous studies have analyzed the biomechanics of nonaxial loads on pedicle screws, but their effects on the screw loosening mechanisms are still not well understood.

METHODS

A design of experiment (DOE) approach was used to evaluate 2 screw thread designs (single-lead and dual-lead threads), 2 insertion trajectories in the transverse and sagittal planes, and 2 loading directions (lateral and cranial). Pedicle screws were inserted in both pedicles of 12 cadaveric lumbar vertebrae for a total of 24 tests. Four sinewave loading cycles (0-400 N) were applied, orthogonally to the screw axis, at the screw head. The resulting forces, displacements, and rotations of the screws were recorded.

RESULTS

In comparison to the other cycles, the first loading cycle revealed important permanent deformation of the bone (mean permanent displacement of the screw head of 0.79 mm), which gradually accumulated over the following cycles to 1.75 mm on average (plowing effect). The cranial loading direction caused significantly lower (P < 0.05) bone deformation than lateral loading. The dual-lead screw had a significantly higher (P < 0.05) initial stiffness than the single-lead thread screw.

CONCLUSIONS

Nonaxial loads induce screw plowing that lead to bone compacting and subsequent screw loosening or even bone failure, thus reducing the pedicle screw fixation strength. Lateral loads induce greater bone deformation and risks of failure than cranial loads.

LEVEL OF EVIDENCE: N/A.

Для цитирования: Bianco Rohan-Jean, Aubin Carl-Eric, Mac-Thiong Jean-Marc, Wagnac Eric, Arnoux Pierre-Jean. Pedicle Screw Fixation Under Nonaxial Loads: A Cadaveric Study.. Spine. 2016. DOI: 10.1097/brs.0000000000001200.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.