STUDY

DESIGN

This was a prospective, randomized, controlled, multicenter study.

OBJECTIVE

The purpose of this study was to compare outcomes of two lumbar total disc replacements (TDRs) at 5-year follow-up and report results of serum ion level analysis in a subgroup of patients receiving a metal-on-metal implant.

SUMMARY OF BACKGROUND DATA: Lumbar TDR has been compared with fusion in several randomized studies, finding TDR noninferior to fusion and superior on some measures. Receiving less attention has been comparing TDR devices. Additionally, there is concern about metal-on-metal implants, with little data available for spine devices.

METHODS

The study included 204 patients receiving Kineflex-L (investigational) and 190 receiving CHARITÉ (control). Outcome measure included Oswestry Disability Index, visual analog pain scales (VAS), patient satisfaction, neurological status, complications, reoperations, and a composite success score. Radiographic assessment included range of motion, subsidence, and heterotrophic ossification. In 32 investigational patients, serum ion analysis was performed for cobalt and chromium. These values were compared with Medicines and Healthcare Products Regulatory Agency values to merit monitoring total hip replacement patients for potential wear problems.

RESULTS

Mean Oswestry and VAS scores in both groups improved significantly by 6 weeks and remained improved during 5-year follow-up (Oswestry Disability Index, scores in both groups were approximately 60 preoperatively vs. 20 at 2- and 5-year follow-up; P < 0.01; VAS scores improved >50% by 6 weeks and remained significantly improved; P < 0.05). Approximately 11% of both groups underwent reoperation. Radiographic analysis found segmental range of motion decreased at 3 month, then increased through 24 months, and was maintained thereafter. Serum ion level analysis found the greatest mean value at any follow-up point was less than 20% of Medicines and Healthcare Products Regulatory Agency recommended minimum value to merit monitoring hip replacement patients.

CONCLUSION

This prospective, randomized study comparing two TDRs found no significant differences in outcomes during 5-year follow-up. Both provided statistically significant improvements by 6 weeks that were maintained. This results support other studies. Serum ion levels in TDR patients were well below the recommended threshold levels to merit monitoring.

Для цитирования: Guyer Richard D, Pettine Kenneth, Roh Jeffrey S, Dimmig Thomas A, Coric Domagoj, McAfee Paul C, Ohnmeiss Donna D. Five-Year Follow-Up of a Prospective, Randomized Trial Comparing Two Lumbar Total Disc Replacements.. Spine. 2016. DOI: 10.1097/brs.0000000000001168.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.