Constrained cups appear incapable of meeting the demands of revision THA.
Аннотация
BACKGROUND
Failure rates of constrained cups for treating recurrent dislocation in revision THA range from 40% to 100%. Although constrained liners are intended to stabilize the hip by mechanically preventing dislocation, the resulting loss of range of motion may lead to impingement and, ultimately, implant failure.
QUESTIONS/PURPOSES: We therefore documented the mechanisms of failure of constrained acetabular cups in revision THA and determined the type and severity of damage (wear, fracture, and impingement) that occurs in situ.
METHODS
We retrieved 57 constrained components of four different designs at revision THA and examined for the presence of rim impingement, oxidation, cracks within the liner, backside wear, pitting, scratching, abrasion, burnishing, and the presence of embedded particles. Articular wear was calculated from the volume of the concave articular bearing surface, which was measured using the fluid displacement method.
RESULTS
Failure of the locking ring was responsible for 51% of failures, whereas 28% of revisions were the result of acetabular cup loosening, 6% backside wear, and 22% infection. Impingement damage of the rim of the polyethylene liner was seen in all retrievals with moderate or severe damage in 54%. The average volumetric wear rate of the articular surface was 95 mm(3)/year.
CONCLUSIONS
Failure of the locking liner ring and loosening of the acetabular cup are the primary causes of mechanical failure with constrained liners; polyethylene is an inadequate material for restricting motion of the hip to prevent instability. The durability of these devices is unlikely to improve unless the mechanical demands are modified through increased range of motion leading to less frequent rim impingement.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.