MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.urology.2023.01.004

Медицинская библиотека · Urology · 2023

Diagnosis of Testicular Torsion and Differentiation From Other Pathologies Using Near-Infrared Spectroscopy.

Ali Üncü Yiğit, Şekerci Çağrı Akın, Yücel Selçuk, Sircan-Kucuksayan Aslınur, Çam Haydar Kamil, Savaş Murat, Yıldırım Asıf, Ağras Koray, Baykara Mehmet, Canpolat Murat

Клиническое исследованиеСтатьяМногоцентровое исследованиеПеревод в очереди

Аннотация

OBJECTIVE

To develop a near-infrared (NIR) spectroscopy device to diagnose testicular torsion with high sensitivity and specificity. Specifically, we aim to investigate the differentiation between testicular torsion from other pathologies such as orchidoepididymitis, varicocele, and hydrocele.

METHODS

Two LEDs with wavelengths of 660 nm and 940 nm were used as light sources in the device. Each wavelength was sent to the testicle successively, and a photodiode detected back-reflected diffuse light. The ratio of the light intensities of 660 nm and 940 nm was used as a diagnostic parameter. A multi-center clinical trial was performed in 5 different hospitals.

RESULTS

In total, 62 patients in urology clinics with acute testicular pain have been recruited for the study. The developed NIR spectroscopy correctly defined all 8 testicular torsion cases. Besides, 3 orchidoepididymitis, 1 varicocele, and 3 hydrocele cases were correctly distinguished from testicular torsion. Only 1 hydrocele case was misdiagnosed as torsion. The range of the ratio was between 0.14 and 1.16 overall measurements. The ratio varied between 0.14 and 0.3 for the testicle with torsion. The ratio was between 0.49 and 1.16 for the normal testicle and testicle with other pathologies mentioned above.

CONCLUSION

We have chosen the threshold ratio of 0.4 to differentiate between the normal and torsion testis and diagnosed all the torsion cases among all normal and other pathologies. The developed optical device to diagnose testicular torsion is inexpensive, user-friendly, and works based on objective criteria with high sensitivity and specificity in real time.

Для цитирования: Ali Üncü Yiğit, Şekerci Çağrı Akın, Yücel Selçuk, Sircan-Kucuksayan Aslınur, Çam Haydar Kamil, Savaş Murat, Yıldırım Asıf, Ağras Koray, Baykara Mehmet, Canpolat Murat. Diagnosis of Testicular Torsion and Differentiation From Other Pathologies Using Near-Infrared Spectroscopy.. Urology. 2023. DOI: 10.1016/j.urology.2023.01.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.