DOI 10.1016/j.arth.2019.12.012
Computed Tomographic Osteolytic Analysis of a First-Generation Remelted Highly Cross-Linked Polyethylene in Total Hip Arthroplasty-At a Minimum of 15-Year Follow-Up.
Аннотация
BACKGROUND
Low polyethylene wear rate and low incidence of osteolysis have been reported after total hip arthroplasty (THA) using first-generation remelted highly cross-linked polyethylene (HXLPE). However, osteolysis has not been evaluated in long-term studies of these patients (15 years or more). The present study assessed computed tomography (CT) images to determine the incidence of osteolysis associated with HXLPE in THA during at least 15 years of follow-up.
METHODS
We evaluated 105 primary THAs in 84 patients (77 women and 7 men). Mean follow-up was 15.9 years (range, 15-18 years). All THAs used a Longevity HXLPE liner and a 26-mm zirconia femoral head. Areas of osteolysis were identified from CT images using 3D and multiplanar reconstruction views.
RESULTS
CT 3D multiplanar reconstruction images at 15-year follow-up showed no obvious osteolysis in the acetabulum or femur of any patient. No patients developed cup loosening or liner breakage.
CONCLUSION
Our study indicates that first-generation remelted HXLPE liners do not increase the risk of osteolysis during 15-year follow-up and suggests that the wear particles from first-generation remelted HXLPE are less biologically active than those generated by conventional polyethylene devices.