MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.clinbiomech.2003.09.010

Медицинская библиотека · Clinical biomechanics (Bristol, Avon) · 2004

Primary stability of a robodoc implanted anatomical stem versus manual implantation.

Nogler Michael, Polikeit Anne, Wimmer Cornelius, Brückner Andreas, Ferguson Stephen J, Krismer Martin

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To assess the initial stability of anatomical stems implanted in manually broached femoral cavities compared with that assessed in cavities milled with the robodoc system.

DESIGN

The bone-prosthesis interface motion was measured in matched pairs of cadaveric femora to assess the initial stability of anatomical stems implanted with two different implantation techniques.

BACKGROUND

The high costs of surgical robots and the increased perioperative efforts associated with their use can only be justified if measurable benefits for patients can be achieved. Increased initial stability of the stem as an early indicator for better bone ongrowth would be such a benefit.

METHODS

Seven pairs of fresh frozen human cadaveric femora were used. One femur of each pair was randomly assigned to receive the robotic milling method; the other femur underwent manual broaching by an experienced surgeon. Initial micromotions of the anatomical stems were measured during simulated gait cycles with loads of < or =1500 N, and both groups underwent matched-pair analysis. Results. High motion of the prostheses was found for both implantation techniques.

CONCLUSIONS

The robodoc system did not enhance the primary stability of the anatomical prosthesis compared with the manual broaching method.

Для цитирования: Nogler Michael, Polikeit Anne, Wimmer Cornelius, Brückner Andreas, Ferguson Stephen J, Krismer Martin. Primary stability of a robodoc implanted anatomical stem versus manual implantation.. Clinical biomechanics (Bristol, Avon). 2004. DOI: 10.1016/j.clinbiomech.2003.09.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.