DOI 10.1016/j.jneumeth.2023.109853

Медицинская библиотека · Journal of neuroscience methods · 2023

Real-Time Semi-Automated and Automated Voxel Placement using fMRI Targets for Repeated Acquisition Magnetic Resonance Spectroscopy.

Bishop James H, Geoly Andrew, Khan Naushaba, Tischler Claudia, Krueger Ruben, Keshava Poorvi, Amin Heer, Baltusis Laima, Wu Hua, Spiegel David, Williams Nolan, Sacchet Matthew D

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Currently, magnetic resonance spectroscopy (MRS) is dependent on the investigative team to manually prescribe, or demarcate, the desired tissue volume-of-interest. The need for a new method to automate precise voxel placements is warranted to improve the utility and interpretability of MRS data.

NEW METHOD: We propose and validate robust and real-time methods to automate MRS voxel placement using functionally defined coordinates within the prefrontal cortex. Data were collected and analyzed using two independent prospective studies: 1) two independent imaging days with each consisting of a multi-session sandwich design (MRS data only collected on one of the days determined based on scan time) and 2) a longitudinal design. Participants with fibromyalgia syndrome (N = 50) and major depressive disorder (N = 35) underwent neuroimaging. MRS acquisitions were acquired at 3-tesla. Evaluation of the reproducibility of spatial location and tissue segmentation was assessed for: 1) manual, 2) semi-automated, and 3) automated voxel prescription approaches

RESULTS

Variability of voxel grey and white matter tissue composition was reduced using automated placement protocols. Spatially, post- to pre-voxel center-of-gravity distance was reduced and voxel overlap increased significantly across datasets using automated compared to manual procedures COMPARISON WITH EXISTING

METHODS

Manual prescription, the current standard in the field, can produce inconsistent data across repeated acquisitions. Using automated voxel placement, we found reduced variability and more consistent voxel placement across multiple acquisitions

CONCLUSIONS

These results demonstrate the within subject reliability and reproducibility of a method for reducing variability introduced by spatial inconsistencies during MRS acquisitions. The proposed method is a meaningful advance toward improved consistency of MRS data in neuroscience and can be utilized for multi-session and longitudinal studies.

Для цитирования: Bishop James H, Geoly Andrew, Khan Naushaba, Tischler Claudia, Krueger Ruben, Keshava Poorvi, Amin Heer, Baltusis Laima, Wu Hua, Spiegel David, Williams Nolan, Sacchet Matthew D. Real-Time Semi-Automated and Automated Voxel Placement using fMRI Targets for Repeated Acquisition Magnetic Resonance Spectroscopy.. Journal of neuroscience methods. 2023. DOI: 10.1016/j.jneumeth.2023.109853.

Предметные рубрики MeSH