MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jmbbm.2019.103434

Медицинская библиотека · Journal of the mechanical behavior of biomedical materials · 2019

Carbon-fibre-reinforced PEEK: An alternative material for flexion bushings of rotating hinged knee joints?

Schroeder Stefan, Braun Steffen, Mueller Ulrike, Vogel Matthias, Sonntag Robert, Jaeger Sebastian, Kretzer Jan Philippe

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

For prosthetic knee joints of the hinged type, typically polyethylene (PE) flexion bushings are used between axis and femoral component to prevent metallic wear. Nevertheless, PE-wear can lead to periprosthetic osteolysis followed by aseptic loosening of the implant. Based on high creep and wear resistance carbon-fibre-reinforced polyether ether ketones (CFR-PEEKs) could provide an alternative material to ultra-high-molecular-weight polyethylene (UHMWPE) for this bearing type.

METHODS

Flexion bushings of four different materials were investigated (CFR-PEEK pitch fibres, CFR-PEEK PAN fibres, virgin PEEK without carbon fibres and UHMWPE) using a bushing tester. For determination of the polymeric and metallic wear, gravimetric measurements and particle analyses were performed.

RESULTS

The polymeric wear rates of CFR-PEEK PAN (6.657 ± 0.714 mg/106 cycles) and CFR-PEEK pitch (32.085 ± 2.748 mg/106 cycles) were significantly higher compared to the polymeric wear rates of virgin PEEK (0.764 ± 0.283 mg/106 cycles) or UHMWPE (-0.015 ± 0.011 mg/106 cycles) (p < .001). In addition, the metallic wear rates were significantly higher when using bushings made of CFR-PEEK PAN (3.373 ± 0.214 mg/106 cycles) and CFR-PEEK pitch (3.604 ± 0.355 mg/106 cycles), than when using bushings made of virgin PEEK (0.177 ± 0.049 mg/106 cycles) or UHMWPE (0.031 ± 0.011 mg/106 cycles) (p < .001). The mean particle size of the two CFR-PEEK groups was significantly smaller compared to the virgin PEEK and UHMWPE group (p < .001).

CONCLUSION

According to these results, the clinical use of CFR-PEEK in combination with CoCr could lead to higher wear rates and this should be avoided. Virgin PEEK may still be considered as an alternative to UHMWPE to use it for flexion bushings in a hinged knee joint.

Для цитирования: Schroeder Stefan, Braun Steffen, Mueller Ulrike, Vogel Matthias, Sonntag Robert, Jaeger Sebastian, Kretzer Jan Philippe. Carbon-fibre-reinforced PEEK: An alternative material for flexion bushings of rotating hinged knee joints?. Journal of the mechanical behavior of biomedical materials. 2019. DOI: 10.1016/j.jmbbm.2019.103434.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.