MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1186/s13018-024-04905-7

Медицинская библиотека · Journal of orthopaedic surgery and research · 2024

Finite element analysis of the treatment of a minimally invasive approach combined with a novel anatomical locking plate for scapular body fractures.

Guo Zhanpeng, Guo Yue, Wang Yansong, Bi Yunlong, Deng Yu, Cao Yang, Huang Mina

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The minimally invasive approach for the treatment of displaced scapular neck or body fractures has the advantages of less trauma and minimal muscle dissection. In clinical practice, the minimally invasive approach combined with an anatomical locking plate has been used to treat scapular body fractures. In addition, we have made minor modifications to the minimally invasive approach. However, the biomechanical study about the approach combined with an anatomical locking plate in treating scapular body fractures was limited.

METHODS

Finite element analysis (FEA) was used to conduct the biomechanical comparison between the anatomical locking plate (AP model) and reconstructive plate (RP model) in the treatment of scapular body fractures through the modified minimally invasive approach. A healthy male volunteer with no history of scapula or systemic diseases was recruited. High-resolution computed tomography images of his right scapula were obtained. Two scapula models were constructed and analyzed by the software of Mimics 21.0, Geomagic Wrap 2021, SolidWorks 2021, and ANSYS Workbench 2022, respectively.

RESULTS

Through static structural analysis, in terms of equivalent von Mises stress, equivalent elastic strain, and total deformation, the AP model exhibited superior safety characteristics, enhanced flexibility, and anticipated stability compared with the RP model. This was evidenced by lower maximum stress, lower maximum strain and displacement.

CONCLUSION

The minimally invasive approach combined with an anatomical locking plate for scapular body fractures had better biomechanical stability. The study provided a biomechanical basis to guide the clinical treatment of scapular body fractures.

Для цитирования: Guo Zhanpeng, Guo Yue, Wang Yansong, Bi Yunlong, Deng Yu, Cao Yang, Huang Mina. Finite element analysis of the treatment of a minimally invasive approach combined with a novel anatomical locking plate for scapular body fractures.. Journal of orthopaedic surgery and research. 2024. DOI: 10.1186/s13018-024-04905-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.