Preventing Iatrogenic Fibula Fractures Using the Push-Pull Technique: A Biomechanical Comparison of Unicortical Versus Bicortical Post Screws.
СтатьяСравнительное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
McCown Sheldon A, Weatherby Paul J, Morris Randal P, Panchbhavi Vinod K, Hagedorn John C, Weiss William M, Chen Jie
Аннотация
BACKGROUND: Displaced diaphyseal fractures can be reduced using the push-pull technique, wherein a plate is affixed to the distal fragment of the fracture, a post screw is placed proximal to the plate, and a lamina spreader creates distraction. This study evaluated the load to failure and mechanism of failure of bicortical and unicortical post screws during reduction.
MATERIALS AND METHODS: Four matched pairs of cadaver legs were subjected to a 2-cm oblique osteotomy simulating a displaced, oblique diaphyseal fracture. A 6-hole compression plate was affixed to the distal fragment with 2 unicortical locking screws, and a 12-mm unicortical or 20-mm bicortical screw was inserted as a post screw proximal to the plate. A lamina bone spreader was used to exert a distraction force between the plate and the post screw. A mechanical actuator simulated the distraction procedure until failure. Maximum applied load, displacement, and absorbed energy were recorded and compared across unicortical and bicortical groups by paired t tests.
RESULTS: At maximum load, we found statistically significant differences in displacement (P=.003) and energy absorbed (P=.022) between the two groups. All unicortical screws failed through screw toggle and bone cut-out. Bicortical screws failed through bending, with no visible damage to the bone at the screw site.
CONCLUSION: When diaphyseal fractures are significantly shortened and require a greater distraction force to achieve reduction, bicortical screws demonstrate a higher mechanical load to failure and increased bone loss from the screw-removal site. A unicortical post screw may be used if minimal distraction is needed. [Orthopedics. 2024;47(5):308-312.].
MATERIALS AND METHODS: Four matched pairs of cadaver legs were subjected to a 2-cm oblique osteotomy simulating a displaced, oblique diaphyseal fracture. A 6-hole compression plate was affixed to the distal fragment with 2 unicortical locking screws, and a 12-mm unicortical or 20-mm bicortical screw was inserted as a post screw proximal to the plate. A lamina bone spreader was used to exert a distraction force between the plate and the post screw. A mechanical actuator simulated the distraction procedure until failure. Maximum applied load, displacement, and absorbed energy were recorded and compared across unicortical and bicortical groups by paired t tests.
RESULTS: At maximum load, we found statistically significant differences in displacement (P=.003) and energy absorbed (P=.022) between the two groups. All unicortical screws failed through screw toggle and bone cut-out. Bicortical screws failed through bending, with no visible damage to the bone at the screw site.
CONCLUSION: When diaphyseal fractures are significantly shortened and require a greater distraction force to achieve reduction, bicortical screws demonstrate a higher mechanical load to failure and increased bone loss from the screw-removal site. A unicortical post screw may be used if minimal distraction is needed. [Orthopedics. 2024;47(5):308-312.].