Rapid urinary lipoarabinomannan test with laboratory-level sensitivity for tuberculosis detection: a performance evaluation.

Jiang Qisheng, Ramachandraiah Harisha, Duncan Carolyn, Gandra Sumanth, George Ige A, Luan Jingyi, Perez Marcos, Lillis Lorraine, Boyle David S, Crick Scott, Ruhwald Morten, Singamaneni Srikanth

СтатьяОценочное исследованиеПеревод в очереди

Аннотация

UNLABELLED: Developing rapid, non-sputum-based tests is a global priority to improve the diagnosis and timely treatment of tuberculosis (TB). In this study, we evaluated the detection limit, precision, stability, and lot-to-lot variation of a next-generation rapid diagnostic test (RDT), the plasmonic fluor (PF)-enhanced urinary lipoarabinomannan lateral flow assay (PF-LAM). We also assessed the diagnostic performance of PF-LAM, Alere Determine TB LAM Ag (AlereLAM), and electrochemiluminescence LAM assay (EclLAM), a highly sensitive laboratory-based test by comparing with microbiology reference standard (MRS) that includes molecular testing (Xpert MTB/RIF) and culture. Two sub-studies were conducted using banked urine samples. First, a preclinical study involving 399 well-characterized urine samples was conducted to determine assay cutoff and assess PF-LAM performance compared with Alere LAM. Second, a diagnostic accuracy assessment study was conducted using 77 blinded samples to compare the performance of PF-LAM to EclLAM. In the preclinical study, when compared with MRS, the AlereLAM showed a sensitivity of 5.5% (95% CI: 3.1%-9.6%) and a specificity of 98.5% (95% CI: 95.6%-99.6%), while PF-LAM exhibited a sensitivity of 33.8% (95% CI: 27.4%-40.6%) and the same specificity of 98.5% (95% CI: 95.6%-99.6%). In the diagnostic accuracy study, when compared with MRS, PF-LAM achieved a sensitivity of 58% (95% CI: 41%-73%) and a specificity of 98% (95% CI: 88%-100%), while EclLAM showed a sensitivity of 42% (95% CI: 27%-59%) and a specificity of 95% (95% CI: 85%-99%). Overall, PF-LAM demonstrated better analytical and diagnostic sensitivity compared with EclLAM and significantly better performance compared with AlereLAM, while exhibiting high precision and minimal lot-to-lot variability.

IMPORTANCE

To the best of our knowledge, this study uniquely reports a rapid diagnostic test, plasmonic fluor-enhanced urinary lipoarabinomannan lateral flow assay (PF-LAM), outperforming an electrochemiluminescence-based laboratory assay. PF-LAM demonstrated promising analytical and diagnostic performance with minimal lot-to-lot variability, positioning it as a helpful tool for non-sputum tuberculosis diagnostics.

Для цитирования: Jiang Qisheng, Ramachandraiah Harisha, Duncan Carolyn, Gandra Sumanth, George Ige A, Luan Jingyi, Perez Marcos, Lillis Lorraine, Boyle David S, Crick Scott, Ruhwald Morten, Singamaneni Srikanth. Rapid urinary lipoarabinomannan test with laboratory-level sensitivity for tuberculosis detection: a performance evaluation.. Microbiology spectrum. 2025. DOI: 10.1128/spectrum.03042-25.